Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-164
2470-901-01 Rev. A /
May
2019
smu.source.output
This attribute enables or disables the source output.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Source configuration list
Function change
Configuration script
Source configuration list
smu.OFF
Usage
sourceOutput
= smu.source.output
smu.source.output =
sourceOutput
sourceOutput
Switch the source output off:
smu.OFF
Switch the source output on:
smu.ON
Details
When the output is switched on, the instrument sources either voltage or current, as set by
smu.source.func
.
Example
smu.source.output = smu.ON
Switch the source output of the instrument to
on.
Also see
Turn the 2470 output on or off
(on page 3-4)
(on page 14-160)
smu.source.protect.level
This attribute sets the overvoltage protection setting of the source output.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Source configuration list
Function change
Configuration script
Source configuration list
smu.PROTECT_NONE
Usage
smu.PROTECT_
x
= smu.source.protect.level
smu.source.protect.level = smu.PROTECT_
x
x
The overvoltage protection level:
20V
,
40V
,
100V
,
200V
,
300V
,
400V
,
500V
, or
NONE
Details
Overvoltage protection restricts the maximum voltage level that the instrument can source. It is in
effect when either current or voltage is sourced.