Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-104
2470-901-01 Rev. A /
May
2019
reset()
This function resets commands to their default settings and clears the buffers.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
reset()
reset(
system
)
system
If the node is the master, the entire system is reset:
true
Only the local group is reset:
false
Details
The
reset()
command in its simplest form resets the entire TSP-enabled system, including the
controlling node and all subordinate nodes.
If you want to reset a specific instrument, use the
node[
N
].reset()
command. Also use the
node[
N
].reset()
command to reset an instrument on a subordinate node.
When no value is specified for
system
, the default value is
true
.
You can only reset the entire system using
reset(true)
if the node is the master. If the node is not
the master node, executing this command generates an error event.
Example
reset(true)
If the node is the master node, the entire system is reset; if the
node is not the master node, an error event is generated.
Also see
(on page 3-49)
script.delete()
This function deletes a script from the run-time memory and nonvolatile memory.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
script.delete("
scriptName
")
scriptName
A string that represents the name of the script
Details
When a script is deleted, the global variable referring to this script is also deleted.
You must delete an existing script before you can use the name of that script again. Scripts are not
automatically overwritten.