2470 High Voltage SourceMeter Instrument Reference Manual
Section 3: Instrument description
2470-901-01 Rev. A /
May
2019
3-49
If an
autoexec
script already exists, you must delete it by sending the
script.delete("autoexec")
command. Performing a system reset does not delete the
autoexec
script.
Resets
There are several types of resets in the 2470.
In general, the terms "reset," "instrument reset," and "system reset" refer to the reset that is
performed when you send the
*RST
or
reset()
command, or when you select
MENU > System >
Info/Manage > System Reset
from the front panel. It resets most commands to their default values.
For more information about the settings that get reset, refer to
(on page 5-28).
The instrument also responds to other types of resets. These resets include:
•
SMU reset:
This reset is only available if you are using the TSP command set. The
smu.reset()
function turns off the output and resets any commands that begin with
smu.
to their default
values. Refer to
(on page 14-150).
•
Password reset:
This resets the instrument password to its default value. You can reset the
password by pressing the
MENU
key, selecting
Info/Manage
(under System), and selecting
Password Reset
. When you do this, the password returns to the default setting. Refer to
(on page 2-37).
•
Digital line reset:
This resets digital I/O line values to their factory defaults if you are using the
TSP command set. If you are using SCPI, the lines are reset when the system is reset.
•
LAN reset:
This resets the LAN settings and the instrument password to the system default
values. To do this reset, insert a straightened paper clip into hole below LAN RESET on the rear
panel. For the location of LAN RESET, refer to
(on page 3-3).
•
Status preset:
This resets all bits in the status model. If you are using the SCPI command set,
refer to
(on page 12-100). If you are using the TSP command set, refer to
(on page 14-184).
•
Trigger blender reset:
This reset is only available if you are using the TSP command set.
Resets some of the trigger blender settings to their factory defaults. Refer to
(on page 14-194).
•
Trigger timer reset:
This reset is only available if you are using the TSP command set. Resets
trigger timer settings to their default values. Refer to
(on page 14-256).
•
TSP-Link line reset:
This reset is only applicable if you are using TSP-Link. Resets some of the
TSP-Link trigger attributes to their defaults. Refer to
(on page 14-272).
•
TSP-Net reset:
This reset is only applicable if you are using TSP-NET. Disconnects all TSP-Net
sessions. Refer to
(on page 14-283).