Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-272
2470-901-01 Rev. A /
May
2019
tsplink.line[N].reset()
This function resets some of the TSP-Link trigger attributes to their factory defaults.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
tsplink.line[
N
].reset()
N
The trigger line (1 to 3)
Details
The
tsplink.line[
N
].reset()
function resets the following attributes to their default values:
•
tsplink.line[
N
].mode
•
trigger.tsplinkin[
N
].edge
•
trigger.tsplinkout[
N
].logic
•
trigger.tsplinkout[
N
].pulsewidth
•
trigger.tsplinkout[
N
].stimulus
This also clears
trigger.tsplinkin[
N
].overrun
.
Example
tsplink.line[3].reset()
Resets TSP-Link trigger line 3 attributes to
default values.
Also see
(on page 14-262)
(on page 14-263)
(on page 14-265)
trigger.tsplinkout[N].pulsewidth
(on page 14-266)
trigger.tsplinkout[N].stimulus
(on page 14-267)
(on page 14-271)