Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-92
2470-901-01 Rev. A /
May
2019
localnode.prompts4882
This attribute enables and disables the generation of prompts for IEEE Std 488.2 common commands.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
No
Power cycle
Not saved
localnode.ENABLE
Usage
prompting
= localnode.prompts4882
localnode.prompts4882 =
prompting
prompting
IEEE Std 488.2 prompting mode:
Disable prompting:
localnode.DISABLE
Enable prompting:
localnode.ENABLE
Details
When this attribute is enabled, the IEEE Std 488.2 common commands generate prompts if
prompting is enabled with the
localnode.prompts
attribute. If
localnode.prompts4882
is
enabled, limit the number of
*trg
commands sent to a running script to 50 regardless of the setting
of the
localnode.prompts
attribute.
When this attribute is disabled, IEEE Std 488.2 common commands will not generate prompts. When
using the
*trg
command with a script that executes
trigger.wait()
repeatedly, disable
prompting to avoid problems associated with the command interface input queue filling.
Example
localnode.prompts4882 = localnode.DISABLE
Disables IEEE Std 488.2 common command
prompting.
Also see
localnode.serialno
This attribute stores the instrument's serial number.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
serialno
= localnode.serialno
serialno
The serial number of the instrument
Details
This indicates the instrument serial number.