2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-125
:TRACe:FILL:MODE
This command determines if a reading buffer is filled continuously or is filled once and stops.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
defbuffer1: CONT
defbuffer2: CONT
User-defined buffers: ONCE
Usage
:TRACe:FILL:MODE <fillType>
:TRACe:FILL:MODE <fillType>, "<bufferName>"
:TRACe:FILL:MODE?
:TRACe:FILL:MODE? "<bufferName>"
<fillType>
Fill the buffer continuously:
CONTinuous
Fill the buffer, then stop:
ONCE
<bufferName>
A string that indicates the reading buffer; the default buffers (
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1
is used
Details
When a reading buffer is set to fill once, no data is overwritten in the buffer. When the buffer is filled,
no more data is stored in that buffer and new readings are discarded.
When a reading buffer is set to fill continuously, the oldest data is overwritten by the newest data after
the buffer fills.
When you change the fill mode of a buffer, any data in the buffer is cleared.
Example
TRACe:MAKE "testData", 100
TRACe:FILL:MODE? "testData"
TRACe:FILL:MODE CONT, "testData"
TRACe:FILL:MODE? "testData"
TRACe:FILL:MODE?
Create a user-defined reading buffer named
testData
with a capacity of 100 measurements.
Query the fill mode setting for
testData
.
Output:
ONCE
Set
testData
fill mode to continuous.
Query the fill mode setting for
testData
.
Output:
CONT
Query the fill mode setting for
defbuffer1
.
Output:
CONT
Also see
(on page 6-25)
(on page 12-127)
(on page 12-119)