2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-75
:SOURce[1]:<function>:DELay:USER<n>
This command sets a user-defined delay that you can use in the trigger model.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Source configuration list
Save settings
Source configuration list
0
Usage
:SOURce[1]:<function>:DELay:USER<n> <delayTime>
:SOURce[1]:<function>:DELay:USER<n> <DEF|MIN|MAX>
:SOURce[1]:<function>:DELay:USER<n>?
<function>
The source function to which this setting applies:
Current:
CURRent
Voltage:
VOLTage
<n>
The number that identifies this user delay (1 to 5)
<delayTime>
The time of the delay in seconds (0 to 10,000)
<DEF|MIN|MAX>
The
DEFault
,
MINimum
, or
MAXimum
value
Details
To use this command in a trigger model, assign the delay to the dynamic delay block.
The delay is specific to the selected function.
Example
:SOUR:VOLT:DEL:USER1 5
:TRIG:BLOC:SOUR:STAT 1, ON
:TRIG:BLOC:DEL:DYN 2, SOUR1
:TRIG:BLOC:MEAS 3
:TRIG:BLOC:SOUR:STAT 4, OFF
:TRIG:BLOC:BRAN:COUN 5, 10, 1
:INIT
Set user delay for source 1 to 5 s.
Set trigger block 1 to turn the source output on.
Set trigger block 2 to a dynamic delay that calls
source user delay 1.
Set trigger block 3 to make a measurement.
Set trigger block 4 to turn the source output off.
Set trigger block 5 to branch to block 1 ten times.
Start the trigger model.
Also see
(on page 12-162)