Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-68
2470-901-01 Rev. A /
May
2019
SOURce subsystem
The commands in the SOURce subsystem configure and control the current source and voltage
source.
:SOURce[1]:CONFiguration:LIST:CATalog?
This command returns the name of one source configuration list.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:SOURce[1]:CONFiguration:LIST:CATalog?
Details
You can use this command to see the names of source configuration lists stored on the instrument.
This command returns one name each time you send it. This command returns an empty string if
there are no more names to return. If the command returns an empty string the first time you send it,
no source configuration lists have been created for the instrument.
Example
:SOUR:CONF:LIST:CAT?
Send this command to return the name of one source configuration
list stored on the instrument. To get all stored configuration lists,
resend this command until it returns an empty string.
Also see
:SOURce[1]:CONFiguration:LIST:CREate
(on page 12-68)
:SOURce[1]:CONFiguration:LIST:CREate
This command creates an empty source configuration list.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:SOURce[1]:CONFiguration:LIST:CREate "<name>"
<name>
A string that represents the name of a source configuration list
Details
This command creates an empty configuration list. To add configuration indexes to this list, you need
to use the store command.