S46T Microwave Switch System High Performance Relay Installation Guide 

 

 

PA-913 Rev. B / December 2019 

 

Figure 5: Relay A through D installation 

 

2.  Refer to the following figure to connect and solder the supplied ribbon cable to the following relays: 

XFER relays (transfer relays); 26.5 GHz unterminated relays; MSPDT relays 

3.  For the A through D relays, secure the ribbon cable to the relay with the two supplied screws. 
4.  Using the following figure, install the relay in the mounting hole, and then secure it with the four screws 

removed in step 1. 

5.  Plug the ribbon connector into the appropriate circuit board connector (see previous figure). 
6.  Install the top cover, and then configure the system as covered in (see 

System configuration

 (on page 9)). 

Summary of Contents for Keithley S46T

Page 1: ...f improper parts may expose the operator to hazardous voltages that could result in personal injury or death Relay kit model number Description Relay location Kit parts summary Quantity Keithley part number Installation reference S46T SPDT KIT 18 GHz SPDT 1 to 8 SPDT unterminated relay 1 SPDT spacer 1 Connecting cable 1 S40 0154 S40 SE05 307 SYS46 311C SPDT relay 1 8 installation on page 3 S46T SP...

Page 2: ...EM RL 318 S40 SV11 350 Relay A D installation on page 6 S46T SP4T KIT 26T 26 5 GHz SP4T A to D 4 40 x 3 16 PPHSEM screws 2 SP4T terminated relay 1 Ribbon cable 1 4 40 x 3 16 PPHSEM RL 310 S40 SV11 350 Relay A D installation on page 6 S46T SP6T KIT 26T 26 5 GHz SP6T A to D 4 40 x 3 16 PPHSEM screws 2 SP6T terminated relay 1 Ribbon cable 1 4 40 x 3 16 PPHSEM RL 290 1 S40 SV11 350 Relay A D installat...

Page 3: ...Reference the following figure and remove the six screws that secure the SPDT relay mounting brackets to the S46T chassis 2 Remove the SPDT assembly from the front panel Figure 1 SPDT relay mounting bracket 3 Reference the following figure and remove the two screws holding the relay spacers in the relay column containing the desired relay installation location 4 Loosen the remaining four SPDT scre...

Page 4: ...on the right side of the instrument column Figure 2 SPDT 1 through 8 relay installation 6 Install the two screws removed in step 2 in the new relay column 7 Thread the red and black wires from the relay through the PVC Tubing PVC tubing TX 52 7 are part of SYS46 311C not shown The TX 52 7 may be cut to a length that will allow the relay assembly to be removed without removing the top cover 8 Refer...

Page 5: ...bracket using the screws removed in step 1 10 Refer to the following table and figure in order to plug the wire connector into the appropriate circuit board connector 11 Install the top cover and configure the system as indicated in System configuration on page 9 Relay Connector Relay Connector Relay Connector 1 J18 5 J22 A J5 2 J19 6 J23 B J6 3 J20 7 J24 C J7 4 J21 8 J25 D J8 ...

Page 6: ... 6 PA 913 Rev B December 2019 Figure 4 Relay connector locations Relay A through D installation To install A through D relays 1 Refer to the following figure and remove the four screws securing the cover plate over the mounting hole where the new relay is to be installed ...

Page 7: ... relays 26 5 GHz unterminated relays MSPDT relays 3 For the A through D relays secure the ribbon cable to the relay with the two supplied screws 4 Using the following figure install the relay in the mounting hole and then secure it with the four screws removed in step 1 5 Plug the ribbon connector into the appropriate circuit board connector see previous figure 6 Install the top cover and then con...

Page 8: ...S46T Microwave Switch System High Performance Relay Installation Guide 8 PA 913 Rev B December 2019 Figure 6 Connect control wires A through D relays ...

Page 9: ...of 0 indicates that a location is not populated For example if RELAY A and RELAY B are SP6T RELAY 1 and RELAY 2 are SPDTs and all other locations are not populated the command syntax would be ROUT CONF CPOL 6 6 0 0 1 1 0 0 0 0 0 0 S46T MSPDT KIT and S46T XFER KIT are programmed with a value of 3 not the number of throws as other relays The MSPDT top relay uses the first channel assignment for rela...

Page 10: ... may be connected to mains These instruments will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to mains Exercise extreme caution when a shock hazard is present Lethal voltage may be present on cable connector jacks or test fixtures The American National Standards Institute ANSI states that a ...

Page 11: ...ocumentation explains hazards that could damage the instrument Such damage may invalidate the warranty The CAUTION heading with the symbol in the user documentation explains hazards that could result in moderate or minor injury or damage the instrument Always read the associated information very carefully before performing the indicated procedure Damage to the instrument may invalidate the warrant...

Reviews: