background image

Model 8020-STC 

 

071321400 / October 2014 

 

Enabling transient overcurrent protection 

Each connector card contains a 10 A fuse to allow maximum pulse current from each SMU instrument. 
Removing this fuse enables transient overcurrent protection. This transient overcurrent protection is useful for 
protecting probes and other sensitive interface hardware at the device under test. For more information, see the 
Model 8020 High Power Interface Panel User's Manual. 

Transient overcurrent protection is only available on the 200 V channels within the Model 8020. When using a 
connector card in the HV channel, be sure that the 10 A fuse is installed. 

 

Summary of Contents for Keithley 8020-STC

Page 1: ...nector Card provides standard triaxial connections for sense and force connections on the Model 8020 High Power Interface Panel Use this connector card to connect to a device under test using standard...

Page 2: ...Model 8020 High Power Interface Panel User s Manual for connector card installation instructions Once you have the card installed and the Model 8020 top cover is reattached and secured you will have...

Page 3: ...rent fuse installed 1 5 A maximum 10 A pulsed 1 ms at 2 2 percent duty cycle Connector current fuse removed 1 ADC no pulsing Connector types 3 lug triaxial Fuse 10 A 32 V MINI blade style Figure 5 Con...

Page 4: ...C 4 071321400 October 2014 Dimensions Figure 7 8020 STC dimensions Installation See the Model 8020 User s Manual for installation information available for download on the Keithley Instruments website...

Page 5: ...vercurrent protection This transient overcurrent protection is useful for protecting probes and other sensitive interface hardware at the device under test For more information see the Model 8020 High...

Page 6: ...be connected to mains These instruments will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to...

Page 7: ...y disposed of according to federal state and local laws The WARNING heading in the user documentation explains dangers that might result in personal injury or death Always read the associated informat...