Section 4: Use the RPM to switch the SMU, CVU, and PMU
Model 4200A-SCS Parameter Analyzer User's Manual
4-2
4200A-900-01 Rev. C / January 2019
Equipment required
One 4200A-SCS with the following instruments:
Two 4200-SMUs or 4210-SMUs
Two 4200-PAs
One 4210-CVU
One 4225-PMU
Two 4225-RPMs
Four 4200-TRX-2 or 4200-MTRX-2 triaxial cables (supplied with SMU)
Four CA-447A SMA cables (supplied with CVU)
Two CA-547-2A RPM interconnect cables (supplied with RPM)
Four CA-534-24A triaxial cables
Two 237-TRX-T triaxial tees
One shielded test fixture with connection to GNDU (Force LO)
Update the RPM configuration in KCon
The KCon application is used to manage the configuration of the 4200A-SCS, including the
4225-RPM. Before using an RPM for automatic switching, you must update the RPM configuration in
KCon. This associates the instruments connected to each RPM and enables automatic switching
between tests.
To update the RPM configuration in KCon:
1.
Make sure your device under test is disconnected from the RPM output terminals.
2.
Close the Clarius application.
3.
Open the KCon application.
4.
Select
Update Preamp, RPM, and CVIV Configuration
.
Figure 29: Update the RPM configuration in KCon
5.
Select
Save
.
6.
Close
KCon
.
7.
Open
Clarius
.