Connect
3.
On the 2601B-P-INT terminal strip panel, use
needle-
nose pliers to position the INTERLOCK jumper.
Place it in the ENABLE slot if you are using an interlock
or in the DISABLE slot if you are not using the interlock.
4.
Align the terminal strip panel of the 2601B-P-INT to the
terminal strips on the rear panel of the 2601B-PULSE, as
shown in the following figure.
5.
Press the 2601B-P-INT connections firmly onto the
terminal strips on the rear panel of the 2601B-PULSE.
6. Use the two 6-32 × ½ in. screws provided with the
2601B-P-INT to secure it to the rear panel of the
2601B-PULSE.
Wiring the interlock
You can use the 2601B-P-INT and the 2601B-PULSE
digital I/O port or external wiring to make interlock
connections to an external device. When properly connected,
the output of the instrument turns OFF when a remote switch
is closed.
Refer to the
Model 2601B-PULSE User’s Manual
(document
number 2601B-PULSE-900-01) for detail on making interlock
connections.