Section 4: Operation
Series 2600B System SourceMeter® instrument User's Manual
4-34
2600BS-900-01 Rev. A / August 2021
To set the current limit in NORMAL output-off mode remotely:*
smu
X
.source.offlimiti =
iValue
To set the voltage limit in NORMAL output-off mode remotely:*
smu
X
.source.offlimitv =
vValue
*
smu
X
can be
smua
for channel A or
smub
for channel B
Remote programming output-off states quick reference
The content of the following table is a quick reference of commands for programming output-off states
from a remote interface.
Output-off state programming quick reference
Command*
Description
smu
X
.source.offmode = smua.OUTPUT_NORMAL
Selects normal output-off mode.
smu
X
.source.offmode = smua.OUTPUT_HIGH_Z
Selects high-impedance output-off mode.
smu
X
.source.offmode = smua.OUTPUT_ZERO
Selects zero output-off mode.
smu
X
.source.offfunc = smua.OUTPUT_DCVOLTS
Sets 0 V output with current limit specified by
the
smua.source.offlimiti
attribute.
smu
X
.source.offfunc = smua.OUTPUT_DCAMPS
Sets 0 A output with voltage limit specified by
the
smua.source.offlimitv
attribute.
smu
X
.source.offlimiti =
iValue
Sets current limit in normal output-off mode.
smu
X
.source.offlimitv =
vValue
Sets voltage limit in normal output-off mode.
*
smu
X
can be
smua
for channel A or
smub
for channel B