Safety precautions 

The following safety precautions should be observed before using this product and any associated instrumentation. Although 

some instruments and accessories would normally be used with nonhazardous voltages, there are situations where hazardous 

conditions may be present. 
This product is intended for use by personnel who recognize shock hazards and are familiar with the safety precautions required 

to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the 

product. Refer to the user documentation for complete product specifications. 

 

If the product is used in a manner not specified, the protection provided by the product warranty may be impaired. 

The types of product users are: 

Responsible body

 is the individual or group responsible for the use and maintenance of equipment, for ensuring that the 

equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained. 

Operators

 use the product for its intended function. They must be trained in electrical safety procedures and proper use of the 

instrument. They must be protected from electric shock and contact with hazardous live circuits. 

Maintenance personnel

 perform routine procedures on the product to keep it operating properly, for example, setting the line 

voltage or replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures 

explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel. 

Service personnel

 are trained to work on live circuits, perform safe installations, and repair products. Only properly trained 

 

service personnel may perform installation and service procedures. 

Keithley

 products are designed for use with electrical signals that are measurement, control, and data I/O connections, with low 

transient overvoltages, and must not be directly connected to mains voltage or to voltage sources with high transient 

overvoltages. Measurement Category II (as referenced in IEC 60664) connections require protection for high transient 

overvoltages often associated with local AC mains connections. Certain 

Keithley

 measuring instruments may be connected to 

mains. These instruments will be marked as category II or higher. 
Unless explicitly allowed in the specifications, operating manual, and instrument labels, do not connect any instrument to mains. 
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test 

fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 

30 V RMS, 42.4 V peak, or 60 VDC are present. A good safety practice is to expect that hazardous voltage is present in any 

unknown circuit before measuring. 
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators 

are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential 

human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If 

 

the circuit is capable of operating at or above 1000 V, no conductive part of the circuit may be exposed. 

Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance-limited 

sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective 

devices to limit fault current and voltage to the card. 
Before operating an instrument, ensure that the line cord is connected to a properly-grounded power receptacle. Inspect the 

connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use. 
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input 

 

power disconnect device must be provided in close proximity to the equipment and within easy reach of the operator. 

For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under 

test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting 

 

cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers. 

Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) 

ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the 

voltage being measured. 