Front panel operation 17
Figure 8
Pulsed waveform
NOTE
The pulse waveform contained in Figure 8 shows trigger level and the integration
time used by the Model 2306 for pulse current measurements.
Front panel operation for performing long integration measurements
All of the following settings are available in sub-menus of the LONG INTEGRAT #1/#2 main
menu item. Scroll through the main menu and when LONG INTEGRAT #1/#2 is displayed
press “ENTER” to access the sub-menus.
Turn on the output
Press the OPERATE key after setting the appropriate voltage and current limits for the DUT.
Select CURRENT RANGE
For Models 2306-PJ and 2302-PJ, select 500mA or 5A. For Models 2306 and 2302, the 5A
range is automatically selected.
Set trigger level and range
“TRIG LEVEL RANGE” and “TRIGGER LEVEL” sub-menus
Battery Channel (#1) — On the 5A current range, the trigger level can be set for either the
5A, 1A, or 100mA range.
Trigger level ranges: 5A range: 0–5A in 5mA steps
1A range: 0–1A in 1mA steps
100mA range: 0–100mA in 0.1 mA steps
“TRG LEV mA RANGE” and “TRIG LEVEL mA” sub-menus
Battery Channel (#1) (Models 2306-PJ and 2302-PJ) — On the 500mA current range, the
trigger level can be set for either the 500mA, 100mA, or the 10mA range.
Trigger level ranges: 500mA range: 0–500mA in 0.5mA steps
100mA range: 0–100mA in 0.1mA steps
10mA range: 0–10mA in 0.1mA steps
Trigger
level
Average current
Integration time
(trigger set to rising)
Integration time
(trigger set to falling)