Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-64
077114601 / March 2019
:BATTery[n]:TEST[:SOURce]:VOLTage:LIMit
This command sets the maximum output voltage in the battery test function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
20 V
Usage
:BATTery[n]:TEST:[SOURce]:VOLTage:LIMit <value>
:BATTery[n]:TEST:[SOURce]:VOLTage:LIMit?
:BATTery[n]:TEST:[SOURce]:VOLTage:LIMit? DEFault
:BATTery[n]:TEST:[SOURce]:VOLTage:LIMit? MINimum
:BATTery[n]:TEST:[SOURce]:VOLTage:LIMit? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<value>
0 to 20 V
Details
This command sets the maximum output voltage in the battery test function.
When this value is set lower than the V-set, the V-set changes to the same value.
Example
:BATT:TEST:VOLT:LIM 5
Set the limit voltage value to 5 V.
Also see
None