Commands Listed in Alphabetical Order
Syntax
[SOURce:]VOLTage[:LEVel][IMMediate][:AMPLitude]
{<voltage>|MIN|MAX}
[SOURce:]VOLTage[:LEVel][IMMediate][:AMPLitude]?
Related Commands
[SOURce:]VOLTage[:LEVel]:DOWN[:IMMediate][:AMPLitude]
[SOURce:]VOLTage[:LEVel]:UP[:IMMediate][:AMPLitude]
[SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]
Arguments
<voltage>
is a
fl
exible decimal number (NRf) that may be type NR1, NR2 or
NR3. It speci
fi
es the voltage setting, which can range from 0 to the maximum
nameplate voltage of the power supply.
MIN
sets the voltage to the minimum level (0 V).
MAX
sets the voltage to the maximum level (note that the maximum level may be
somewhat higher than the nameplate).
UP
sets the voltage level to increase a step.
DOWN
sets the voltage level to decrease a step.
DEF
is the default level (1 V).
Returns
NR2
is the voltage setting in volts.
Examples
VOLTAGE:MIN
VOLTAGE?
might return
1.05
, which would be the voltage setting in volts.
[SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]
This command is used to set the voltage step value.
Group
Source
Syntax
[SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]
{<voltage>}
[SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]?
Related Commands
[SOURce:]VOLTage[:LEVel]:DOWN[:IMMediate][:AMPLitude]
Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual