Section 6: Current function adjustment (DC and AC)
Model 2100 6½-Digit Resolution Digital Multimeter
6-2
2100-905-01 Rev. B April 2020
12. Set the 2100 to the 0.1 A dc current range by sending the following SCPI command:
CONF:CURR:DC 0.1
13. Set the measurement rate to 10 PLC by sending the following SCPI command:
CURR:NPLC 10
14. Set the 5700A to 0 A and output mode to
OPERATE
.
15. Adjust the front input offset by sending the following SCPI command:
CAL:PROT:DC:STEP 1,0.1
16. Wait until the following message is displayed on the 2100:
Cali OK
17. Set the 5700A output to 0.06 A dc.
18. Adjust the mid-region of the range by sending the following SCPI command:
CAL:PROT:DC:STEP 2,0.06
19. Wait until the following message is displayed on the 2100:
Cali OK
20. Set the 5700A output to 0.12 A dc.
21. Adjust the full-scale measurement by sending the following SCPI command:
CAL:PROT:DC:STEP 2,0.12
22. Wait until the following message is displayed on the 2100:
Cali OK
23. Set the 5700A output mode to
STANDBY
.
24. Set the 2100 to the 1 A dc current range by sending the following SCPI command:
CONF:CURR:DC 1
25. Set the measurement rate to 10 PLC by sending the following SCPI command:
CURR:NPLC 10
26. Set the 5700A to 0 A and output mode to
OPERATE
.
27. Adjust the front input offset by sending the following SCPI command:
CAL:PROT:DC:STEP 1,1
28. Wait until the following message is displayed on the 2100:
Cali OK
29. Set the 5700A output to 1 A dc.
30. Adjust the full-scale measurement by sending the following SCPI command:
CAL:PROT:DC:STEP 2,1
31. Wait until the following message is displayed on the 2100:
Cali OK
32. Set the 5700A output mode to
STANDBY
. Press the
RESET
button on 5700A front panel to reset
the calibrator to its default settings.