Operating information
For example, the IVTIP5X tip has a dynamic range of 5 V
p-p
when the 1X Range
is selected. This means that you can display a signal with a ±2.5 V differential
voltage. When the 2X Range is selected, the dynamic range increases from 5 V
p-p
to 10 V
p-p
(±5.0 V). Refer to the
Linear differential input voltage range
in the
speci
fi
cations table for more information.
CAUTION.
To avoid damaging the probe, do NOT exceed either the RMS OR
Peak Voltage ratings. The Maximum Non-Destruct Voltage limits (RMS and Peak
Voltage) do not increase when the 2X Range is selected. For the IVTIP5X, the
12 V
rms
and ±21.5 V
pk
limits are the same for the 1X or 2X Range.
Auto Range
By default, the MSO/DPO5000, DPO7000, and MSO/DPO70000 Series
Oscilloscopes select the 1X or 2X Range automatically when you change the
V/div setting. This hides the complexity of selecting the range from casual users.
However, there are combinations of Range and V/div settings that cannot be
reached when Auto Range is selected. For these situations, select Manual Range
when full
fl
exibility is desired.
18
TIVM Series Measurement System User Manual
Summary of Contents for IsoVu TIVM Series
Page 1: ...xx TIVM Series IsoVu Measurement System ZZZ User Manual P071349500 071 3495 00...
Page 2: ......
Page 6: ......
Page 18: ...Preface xii TIVM Series Measurement System User Manual...
Page 40: ...Operating information 22 TIVM Series Measurement System User Manual...
Page 46: ...Application examples 28 TIVM Series Measurement System User Manual...
Page 72: ...User service 54 TIVM Series Measurement System User Manual...
Page 80: ...Appendix B Compliance information 62 TIVM Series Measurement System User Manual...