Save and Recall Information
Side-bezel menu button
Description
Displaying a reference
waveform
Sets the oscilloscope to save waveform data as a comma-separated data
fi
le compatible with
popular spreadsheet programs. This
fi
le cannot be recalled to reference memory.
Saving an analog waveform to reference memory.
To save an analog waveform to nonvolatile memory inside the
oscilloscope, select the waveform that you want to save, push the
Save Waveform
screen button, and then select one of the
reference waveform locations. Four-channel models have four reference locations. Two-channel models have two reference
locations.
Saved waveforms contain only the most recent acquisition. Gray-scale information, if any, is not saved.
NOTE.
10 M reference waveforms are volatile and not saved when the oscilloscope power is turned off. To keep these waveforms,
save them to external storage.
Displaying a reference waveform.
To display a waveform stored in nonvolatile memory:
1.
Push
Ref R
.
MSO4000 and DPO4000 Series Oscilloscopes User Manual
245