Commands Listed in Alphabetical Order
WFMInpre:YOFf
Sets or returns the vertical position of the incoming waveform in digitizing levels.
Variations in this number are analogous to changing the vertical position of the
waveform.
YMUlt, YOFf, and YZEro are used to convert waveform record values to YUNit
values using the following formula (where dl is the data level; curve_in_dl is a
data point in CURVe):
value_in_units = ((curve_in_dl - YOFf_in_dl) * YMUlt) + YZEro_in_units
NOTE.
For a given waveform record, YMUlt, YOFf, and YZEro have to be a
consistent set, otherwise vertical cursor readouts and vertical measurements may
give incorrect results.
Group
Waveform Transfer
Syntax
WFMInpre:YOFf <NR3>
WFMInpre:YOFf?
Related Commands
,
,
Arguments
<NR3>
is the vertical offset in digitizing levels.
Examples
WFMINPRE:YOFF 50
speci
fi
es that the zero reference point for the incoming
waveform is 50 digitizing levels (2 divisions, for 8-bit data) above the center
of the data range.
WFMINPRE:YOFF?
might return
:WFMINPRE:YOFF 25
indicating the vertical
position of the incoming waveform in digitizing levels.
WFMInpre:YUNit
Sets or returns the vertical units of the incoming waveform.
Supported units are: %, /Hz, A, A/A, A/V, A/W, A/dB, A/s, AA, AW, AdB, As,
B, Hz, IRE, S/s, V, V/A, V/V, V/W, V/dB, V/s, VV, VW, VdB, Volts, Vs, W,
W/A, W/V, W/W, W/dB, W/s, WA, WV, WW, WdB, Ws, dB, dB/A, dB/V, dB/W,
dB/dB, dBA, dBV, dBW, dBdB, day, degrees, div, hr, min, ohms, percent, s
Group
Waveform Transfer
MSO4000 and DPO4000 Series Programmer Manual
2-517
Summary of Contents for DPO4000 Series
Page 2: ......
Page 10: ...Table of Contents vi MSO4000 and DPO4000 Series Programmer Manual ...
Page 23: ...Command Syntax MSO4000 and DPO4000 Series Programmer Manual 2 9 ...
Page 24: ...Command Syntax 2 10 MSO4000 and DPO4000 Series Programmer Manual ...
Page 84: ...Command Groups 2 70 MSO4000 and DPO4000 Series Programmer Manual ...
Page 546: ...Commands Listed in Alphabetical Order 2 532 MSO4000 and DPO4000 Series Programmer Manual ...
Page 564: ...Status and Events 3 18 MSO4000 and DPO4000 Series Programmer Manual ...
Page 565: ...Appendix A Character Set MSO4000 and DPO4000 Series Programmer Manual A 1 ...
Page 566: ...Appendix A Character Set A 2 MSO4000 and DPO4000 Series Programmer Manual ...
Page 572: ...Appendix B Reserved Words B 6 MSO4000 and DPO4000 Series Programmer Manual ...