Commands Listed in Alphabetical Order
DATa:STARt and DATa:STOP are order independent. When DATa:STOP is
greater than DATa:STARt, the values will be swapped internally for the CURVE?
query.
Examples
DATA:START?
might return
:DATA:START 214
indicating that data point 214 is
the
fi
rst waveform data point that will be transferred.
DATA:START 10
speci
fi
es that the waveform transfer will begin with data point 10.
DATa:STOP
Sets or returns the last data point that will be transferred when using the
?
query. This command allows for the transfer of partial waveforms from the
oscilloscope.
Changes to the record length value are not automatically re
fl
ected in the
DATa:STOP
value. As record length is varied, the
DATa:STOP
value must be
explicitly changed to ensure the entire record is transmitted. In other words, curve
results will not automatically and correctly re
fl
ect increases in record length if
the distance from
to
DATa:STOP
stays smaller than the increased
record length.
Group
Waveform Transfer
Syntax
DATa:STOP <NR1>
DATa:STOP?
Related Commands
,
,
,
Arguments
<NR1>
is the last data point that will be transferred, which ranges from 1 to
the record length. If
<NR1>
is greater than the record length, then data will be
transferred up to the record length. If both DATa:STARt and DATa:STOP are
greater than the record length, the last data point in the record is returned.
DATa:STARt and DATa:STOP are order independent. When DATa:STOP is less
than DATa:STARt, the values will be swapped internally for the CURVE? query.
If you always want to transfer complete waveforms, set DATa:STARt to 1 and
DATa:STOP to the maximum record length, or larger.
Examples
DATA:STOP?
might return
:DATA:STOP 14900
indicating that 14900 is the last
waveform data point that will be transferred.
MSO4000 and DPO4000 Series Programmer Manual
2-155
Summary of Contents for DPO4000 Series
Page 2: ......
Page 10: ...Table of Contents vi MSO4000 and DPO4000 Series Programmer Manual ...
Page 23: ...Command Syntax MSO4000 and DPO4000 Series Programmer Manual 2 9 ...
Page 24: ...Command Syntax 2 10 MSO4000 and DPO4000 Series Programmer Manual ...
Page 84: ...Command Groups 2 70 MSO4000 and DPO4000 Series Programmer Manual ...
Page 546: ...Commands Listed in Alphabetical Order 2 532 MSO4000 and DPO4000 Series Programmer Manual ...
Page 564: ...Status and Events 3 18 MSO4000 and DPO4000 Series Programmer Manual ...
Page 565: ...Appendix A Character Set MSO4000 and DPO4000 Series Programmer Manual A 1 ...
Page 566: ...Appendix A Character Set A 2 MSO4000 and DPO4000 Series Programmer Manual ...
Page 572: ...Appendix B Reserved Words B 6 MSO4000 and DPO4000 Series Programmer Manual ...