Commands Listed in Alphabetical Order
NOTE.
Curve data is transferred from the oscilloscope asynchronously, depending
on the length of the curve record. Such transfers may require several seconds
to complete. During this period, the oscilloscope will not respond to the user
controls. You can interrupt these asynchronous data transfers by sending a device
clear message to the oscilloscope or by interrupting the query with another
command or query. In order to verify that curve data has been completely
transferred, it is recommended that you follow such queries with an
query to verify there are no error bits set. You can also check the event queue to
determine the cause of the error. If the error was caused by an interrupted query,
then the asynchronous data transfer was not complete when the
query was
sent. In such cases, it may be necessary to increase the program’s time-out value
to ensure that all data is transferred and read.
DATa
Sets or returns the format and location of waveform data transferred with the
? query or
command.
Group
Waveform Transfer
Syntax
DATa {INIT|SNAp}
DATa?
Related Commands
,
,
Arguments
INIT
initializes the waveform data parameters to their factory defaults except for
, which is set to the current acquisition record length.
SNAp
sets
and
to match the current waveform cursor
positions.
Examples
DATA?
might return
:DATA:DESTINATION REF1:ENCDG
RIBINARY;SOURCE
CH1;START 1;STOP 500;WIDTH 1
DATA INIT
initializes the waveform data parameters to their factory defaults.
DATa:DESTination
Sets or returns the reference memory location for storing waveform data
transferred into the oscilloscope by the
command.
MSO4000 and DPO4000 Series Programmer Manual
2-151
Summary of Contents for DPO4000 Series
Page 2: ......
Page 10: ...Table of Contents vi MSO4000 and DPO4000 Series Programmer Manual ...
Page 23: ...Command Syntax MSO4000 and DPO4000 Series Programmer Manual 2 9 ...
Page 24: ...Command Syntax 2 10 MSO4000 and DPO4000 Series Programmer Manual ...
Page 84: ...Command Groups 2 70 MSO4000 and DPO4000 Series Programmer Manual ...
Page 546: ...Commands Listed in Alphabetical Order 2 532 MSO4000 and DPO4000 Series Programmer Manual ...
Page 564: ...Status and Events 3 18 MSO4000 and DPO4000 Series Programmer Manual ...
Page 565: ...Appendix A Character Set MSO4000 and DPO4000 Series Programmer Manual A 1 ...
Page 566: ...Appendix A Character Set A 2 MSO4000 and DPO4000 Series Programmer Manual ...
Page 572: ...Appendix B Reserved Words B 6 MSO4000 and DPO4000 Series Programmer Manual ...