Commands Listed in Alphabetical Order
When :BYT_Nr is 1, the range is from -128 through 127. When :BYT_Nr is 2,
the range is from -32,768 through 32,767. Center screen is 0 (zero). The upper
limit is the top of the screen and the lower limit is the bottom of the screen. This
is the default argument.
RPBinary
speci
fi
es the positive integer data-point representation, with the most
signi
fi
cant byte transferred
fi
rst.
When :BYT_Nr is 1, the range from 0 through 255. When :BYT_Nr is 2, the
range is from 0 to 65,535. The center of the screen is 127. The upper limit is the
top of the screen and the lower limit is the bottom of the screen.
SRIbinary
is the same as
RIBinary
except that the byte order is swapped,
meaning that the least signi
fi
cant byte is transferred
fi
rst. This format is useful
when transferring data to IBM compatible PCs.
SRPbinary
is the same as
RPBinary
except that the byte order is swapped,
meaning that the least signi
fi
cant byte is transferred
fi
rst. This format is useful
when transferring data to PCs.
Table 2-39: DATa and WFMOutpre Parameter Settings
WFMOutpre Settings
DATa:ENCdg
Setting
:ENCdg
:BN_Fmt
:BYT_Or
:BYT_NR
ASCii
ASC
N/A
N/A
1,2
FAStest
BIN
RI
MSB
1,2
RIBinary
BIN
RI
MSB
1,2
RPBinary
BIN
RP
MSB
1,2
SRIbinary
BIN
RI
LSB
1,2
SRPbinary
BIN
RP
LSB
1,2
Examples
DATA:ENCDG?
might return
:DATa:ENCDG SRPBINARY
for the format of the
outgoing waveform data.
DATA:ENCDG RPBinary
sets the data encoding format to be a positive integer
where the most signi
fi
cant byte is transferred
fi
rst.
DATa:SOUrce
Sets or returns the location of the waveform data transferred from the oscilloscope
by the
? query.
Group
Waveform Transfer
MSO4000 and DPO4000 Series Programmer Manual
2-153
Summary of Contents for DPO4000 Series
Page 2: ......
Page 10: ...Table of Contents vi MSO4000 and DPO4000 Series Programmer Manual ...
Page 23: ...Command Syntax MSO4000 and DPO4000 Series Programmer Manual 2 9 ...
Page 24: ...Command Syntax 2 10 MSO4000 and DPO4000 Series Programmer Manual ...
Page 84: ...Command Groups 2 70 MSO4000 and DPO4000 Series Programmer Manual ...
Page 546: ...Commands Listed in Alphabetical Order 2 532 MSO4000 and DPO4000 Series Programmer Manual ...
Page 564: ...Status and Events 3 18 MSO4000 and DPO4000 Series Programmer Manual ...
Page 565: ...Appendix A Character Set MSO4000 and DPO4000 Series Programmer Manual A 1 ...
Page 566: ...Appendix A Character Set A 2 MSO4000 and DPO4000 Series Programmer Manual ...
Page 572: ...Appendix B Reserved Words B 6 MSO4000 and DPO4000 Series Programmer Manual ...