Reference
CMC251 User Manual
11
Frequencies Between 80 MHz and 1.3 GHz
1. Press the FUNC button until FREQ B is indicated by a lighted
LED.
2. After determining that the signal voltage level is 1 V rms or less,
connect the signal to the CHANNEL B INPUT connector.
3. Press the GATE button to select the desired display resolution.
Refer to Tables 4 and 5 for resolution-versus-gate settings.
Table 4: Gate Time Versus Resolution for Channel A
GATE Setting
Gate Time
Resolution
PRESCALE Off
PRESCALE On
X1
0.01 seconds
100 Hz
1 kHz
X10
0.1 seconds
10 Hz
100 Hz
X100
1 second
1 Hz
10 Hz
X1000
10 seconds
0.1 Hz
1 Hz
Table 5: Gate Time Versus Resolution for Channel B
GATE Setting
Gate Time
Resolution
X1
0.016 seconds
10 kHz
X10
0.16 seconds
1 kHz
X100
1.6 seconds
100 Hz
X1000
16 seconds
10 Hz
Making Period Measurements
A Period measurement is the time in seconds for a signal to complete
one entire cycle. This measurement is taken by averaging the signal
events for a certain number of internal clock pulses. Therefore, the
gate setting determines the resolution (or accuracy) of the reading.
Summary of Contents for CMC251
Page 1: ...User Manual CMC251 1 3 GHz Frequency Counter 070 8527 03...
Page 4: ......
Page 10: ...General Safety Summary vi CMC251 User Manual...
Page 18: ...Getting Started 8 CMC251 User Manual...
Page 28: ...Reference 18 CMC251 User Manual...
Page 34: ...Appendix A Specifications 24 CMC251 User Manual...