![Tektronix CA PLUG-IN Instruction Manual Download Page 18](http://html1.mh-extra.com/html/tektronix/ca-plug-in/ca-plug-in_instruction-manual_1078993018.webp)
7. Tektronix 20pf Input Capacitance Standard-
iz e r , Tektronix part number 011-022.
8. 50 ohm 5XT attenuator, Tektronix Part
Number 011-032.
9. 50 ohm cable termination, Tektronix Part
Number 011-045.
10. Tektronix Type P52 C oaxialC ab les,52ohms
characteristic impedance.
11. Tektronix Dual Input Connector, see Fig.
5-1.
Tektronix
Part Number 003-035.
resistance-to-ground checks at thelö-pininter-
connecting plug.
The table below lists the
nominal resistance value from each pin to
ground.
NOMINAL RESISTANCES AT
INTERCONNECTING PLUG
PIN NUMBER
RESISTANCE-TO-GROUND
1
8.5k
2
0
3
8.5k
4
infinite
5
infinite
6
infinite
7
infinite
8
0 (MODE switch
at ALTERN ATE )
9
12k
10
2k
11
6.3k
12
infinite
13
infinite
14
infinite
15
650
16
infinite
12. Miscellaneous alignment tools. See Fig.
5-2.
PROCEDURE
1. Prelim inary Check
Install the Type CA Unit in the oscilloscope.
Turn on all the test equipment and allow
15 minutes fo r it to warm up. Set the o s c illo
scope Controls as follows:
B efore installing the Type CA Plug-In Unit
in the oscilloscope, make a careful visual
inspection of the w ire dress. This
is par-
ticularly important if any soldering has been
done to the unit. Then make the following
F ig . 5-2. L o w capacitance, insulated alignment tools used to ca lib ra te
the T yp e C A .
HORIZONTAL D ISPLAY
INTERNAL
SWEEP (541A)
MAIN SWEEP
NORMAL
(545 A)
TRIGGERING MODE
TRIGGER SLOPE
S T A B IL IT Y
TIME/CM
AUTOMATIC
-IN T
PRESET
1 MILLISEC
Now preset the Type CA Controls as follows:
VOLTS/CM (A and B)
.05
AA
1
5-2
Calibration - Type CA
Summary of Contents for CA PLUG-IN