Table 1: External inspection checklist
Item
Inspect for
Repair action
Cabinet, front panel, and cover
Cracks, scratches, deformations,
damaged hardware
Repair or replace defective module
Front-panel knobs
Missing, damaged, or loose knobs
Repair or replace missing or defective
knobs
Connectors
Broken shells, cracked insulation, and
deformed contacts. Dirt in connectors
Repair or replace defective modules.
Clear or brush out dirt
Carrying handle and cabinet feet
Correct operation
Repair or replace defective module
Accessories
Missing items or parts of items, bent
pins, broken or frayed cables, and
damaged connectors
Repair or replace damaged or missing
items, frayed cables, and defective
modules
Flat panel display cleaning
Clean the flat panel display surface by gently rubbing the display with a clean-room wipe (such as Wypall Medium Duty
Wipes, #05701, available from Kimberly-Clark Corporation), or an abrasive-free cleaning cloth.
If the display is very dirty, moisten the wipeor cloth with distilled water, a 75% isopropyl alcohol solution, or standard glass
cleaner, and gently rub the display surface. Use only enough liquid to dampen the cloth or wipe. Avoid using excess force or
you may damage the display surface.
CAUTION: Improper cleaning agents or methods can damage the flat panel display.
• Do not use abrasive cleaners or surface cleaners to clean the display.
• Do not spray liquids directly on the display surface.
• Do not scrub the display with excessive force.
CAUTION: To prevent getting moisture inside the instrument during external cleaning, do not spray any cleaning
solutions directly onto the screen or instrument.
Interior cleaning
You can only clean the interior of the rear chassis assembly. To access the rear chassis assembly, see the Removal and
Installation Procedures topics.
Do not disassemble or clean the front chassis assembly (other than cleaning the front panel display and removing panel
knobs). Disassembling the front chassis assembly requires that the instrument be returned to your nearest Tektronix Service
Center for calibration.
Warning: To avoid electric shock or damage to the instrument, remove instrument power. Before performing any
procedure that follows, power down the instrument and disconnect it from line voltage.
• Blow off dust with dry, low-pressure (approximately 9 psi), deionized air.
• Use a soft-bristle, non-static-producing brush for cleaning around components.
• Remove any remaining dust with a lint-free cloth dampened in isopropyl alcohol (75% solution). (A cotton-tipped
applicator is useful for cleaning in narrow spaces and on circuit boards).
• If you must use a liquid for minor rear chassis interior cleaning, use a clean-room wipe (such as Wypall Medium Duty
Wipes, #05701, available from Kimberly-Clark Corporation) dampened with a 75% isopropyl alcohol solution.
Maintenance
6 Series B MSO MSO64B, MSO66B, MSO68B Service
6