Troubleshooting
4-42
AWG7000 Series Service Manual
Checking the Power
Supply Voltages
To check the power supply voltages, power on the instrument and connect the (–)
lead of a multi-meter to chassis ground.
To measure the output voltages of the power supply, check the voltages at J104 on
the PWR board with the multi-meter and compare each reading to the values listed
in Table 4-4. If the voltages are within the allowance, your power supply is
functional.
Figure 4-20: PWR board test points
Table 4-4: Power supply voltages
PWR board (J104)
Voltage
Allowance
Pin1
+3.3 V
+3.1 V to +3.5 V
Pin 3
+5 V
+4.7 V to 5.3 V
Pin 5
+12 V
+11.4 V to +12.6 V
Pin 1
+ 3.3 V
Pin 3
+ 5V
Pin 5
+ 12 V
Multi-meter (+) lead