Troubleshooting
AWG7000 Series Service Manual
4-41
Figure 4-19: Primary troubleshooting tree (2)
Diagnostics
passed?
Amplitude OK?
Frequency OK?
Output
waveform OK?
Check the diagnostics error
code on page 4-45.
Replace the related boards.
Replace the AWG10G or
OUT10G board.
Replace the CLK10G
board.
Replace the AWG10G
board.
No
No
No
No
Yes
Yes
Yes
Yes
Yes
Yes
1
Generator troubleshoot
Execute System menu >
Diagnostics...
Execute System menu >
Calibration...
Calibration
passed?
Check the calibration error
code on page 4-50.
Replace the related boards.
No
Execute functional test
described in Performance
Verification manual.
Completed.