Adjustment Procedures
3-18
AWG7000 Series Service Manual
9.
Identify the minimum amplitude (ripple) point for the Ch1 waveform on the
sampling oscilloscope screen while changing the
Ch2 HFD205 Phase
settings
using either the up/down button or the front-panel rotary knob. You can also
enter the value directly using the keyboard.
The
Skew
for each sampling rate is calculated automatically using the
Inter-Channel Skew Cal. results, and the skew values are displayed in the
Skew5G
through
Skew10G
fields.
10.
Stop at the minimum ripple point, and select
6.0 GHz
in the Sampling Rate.
11.
Repeat step 9.
12.
Stop at the minimum ripple point, and select
7.0 GHz
in the Sampling Rate.
13.
Repeat step 9.
14.
Stop at the minimum ripple point, and select
8.0 GHz
in the Sampling Rate.
15.
Repeat step 9.
16.
Stop at the minimum ripple point, and select
9.0 GHz
in the Sampling Rate.
17.
Repeat step 9.
18.
Stop at the minimum ripple point, and select
10.0 GHz
in the Sampling Rate.
19.
Repeat step 9.
20.
Stop at the minimum ripple point, and click
Update Cal Values
.
21.
Click
Save to EEPROM
.
22.
Click
Close
. The Interleave Calibration window closes.
23.
Click
Exit
in the Service UI menu.
This completes the adjustment procedures. Disconnect the equipment from the
instrument.
NOTE.
Ideally, the skew should not depend on the frequency, but there is a slight
dependency between skew and frequency.
Summary of Contents for AWG7000 Series
Page 4: ......
Page 10: ...List of Tables vi AWG7000 Series Service Manual...
Page 14: ...Service Safety Summary x AWG7000 Series Service Manual...
Page 16: ...Environmental Considerations xii AWG7000 Series Service Manual...
Page 18: ...Preface xiv AWG7000 Series Service Manual...
Page 19: ...Operating Information...
Page 20: ......
Page 23: ...Theory of Operation...
Page 24: ......
Page 31: ...Adjustment Procedures...
Page 32: ......
Page 51: ...Maintenance...
Page 52: ......
Page 58: ...Maintenance 4 6 AWG7000 Series Service Manual...
Page 109: ...Replaceable Parts...
Page 110: ......