Adjustment Procedures
3-12
AWG7000 Series Service Manual
Inter-Channel Skew
Calibration
(AWG7102 and AWG7052)
This calibration is required to ensure that the Ch 1 and Ch 2 output skews are set
within the specified range. This calibration is applied to both analog and marker
outputs.
1.
Install the test hookup and preset the instrument controls:
a.
Hook up the sampling oscilloscope:
Attach the SMA terminations to the Channel 1 Analog (–) Output,
Channel 2 Analog (–) Output, and Marker 1 (–) Output on the front
panel of the arbitrary waveform generator.
Connect an SMA cable from the Channel 1 Analog (+) Output on the
front panel of the arbitrary waveform generator to the CH 1 Input of
the sampling oscilloscope using the SMA attenuator, as shown in
Figure 3-6.
Connect an SMA cable from the Channel 1 Marker (+) Output on the
front panel of the arbitrary waveform generator to the TRIGGER
DIRECT INPUT of the sampling oscilloscope using the SMA
attenuator, as shown in Figure 3-6.
Figure 3-6: Inter-channel skew calibration initial hookup
Equipment required
One sampling oscilloscope
Two 50
Ω
SMA 12 dB attenuators
Two 50
Ω
SMA cables
Three 50
Ω
SMA terminations
SMA termination
Instrument under test
Sampling oscilloscope
SMA attenuator
SMA attenuator
Summary of Contents for AWG7000 Series
Page 4: ......
Page 10: ...List of Tables vi AWG7000 Series Service Manual...
Page 14: ...Service Safety Summary x AWG7000 Series Service Manual...
Page 16: ...Environmental Considerations xii AWG7000 Series Service Manual...
Page 18: ...Preface xiv AWG7000 Series Service Manual...
Page 19: ...Operating Information...
Page 20: ......
Page 23: ...Theory of Operation...
Page 24: ......
Page 31: ...Adjustment Procedures...
Page 32: ......
Page 51: ...Maintenance...
Page 52: ......
Page 58: ...Maintenance 4 6 AWG7000 Series Service Manual...
Page 109: ...Replaceable Parts...
Page 110: ......