Adjustment Procedures
AWG7000 Series Service Manual
3-7
This calibration consists of six steps from Step 0 to Step 5, but in this manual,
perform only Step 4 (Magic Cal.-2). Other calibration constants are restored by
using the data acquired by ATS at the factory or during servicing.
Basic steps are to observe output waveforms on the oscilloscope screen, and to
determine the range of abnormal outputs, which appears on the oscilloscope as a
broken waveform, by adjusting the delay in the TEK0015 Data Output.
1.
Install the test hookup and preset the instrument controls:
a.
Hook up the oscilloscope:
Attach the SMA terminations to the Channel 1 Analog (–) and Channel 2
Analog (–) Outputs on the front panel of the arbitrary waveform generator.
Attach the SMA-BNC adapters to the CH 1 and CH 2 Input connectors of
the oscilloscope.
Connect the SMA cables from the Channel 1 Analog (+) and Channel 2
Analog (+) Outputs on the front panel of the arbitrary waveform generator
to the CH 1 and CH 2 Inputs of the oscilloscope, as shown in Figure 3-4.
Figure 3-4: Data timing calibration initial hookup
Equipment required
One oscilloscope
Two SMA-BNC adapters
Two 50
Ω
SMA cables
Two 50
Ω
SMA terminations
SMA-BNC adapters
Instrument under test
Oscilloscope
SMA terminations
SMA cables
Summary of Contents for AWG7000 Series
Page 4: ......
Page 10: ...List of Tables vi AWG7000 Series Service Manual...
Page 14: ...Service Safety Summary x AWG7000 Series Service Manual...
Page 16: ...Environmental Considerations xii AWG7000 Series Service Manual...
Page 18: ...Preface xiv AWG7000 Series Service Manual...
Page 19: ...Operating Information...
Page 20: ......
Page 23: ...Theory of Operation...
Page 24: ......
Page 31: ...Adjustment Procedures...
Page 32: ......
Page 51: ...Maintenance...
Page 52: ......
Page 58: ...Maintenance 4 6 AWG7000 Series Service Manual...
Page 109: ...Replaceable Parts...
Page 110: ......