Adjustment Procedures
This section contains information about instrument adjustment. Only quali
fi
ed
personnel should perform adjustment procedures. Read the Service Safety
Summary and the General Safety Summary before performing any service
procedures.
NOTE.
Before performing adjustment procedures, you must warm up the arbitrary
waveform generator at least 20 minutes in an ambient temperature between 20 °C
and 30 °C. Adjustments performed before warm-up or outside this temperature
range may result in poor performance.
Adjustment Interval
If the instrument fails performance tests (refer to the
AWG7000 Series Arbitrary
Waveform Generators Technical Reference Manual
), adjustment may be required.
If periodic calibration is one of your requirements, a general rule is to verify
performance and make adjustments (if needed) every 2000 hours of operation (or
once a year if the instrument is used infrequently).
Adjustment After Repair
After removal and replacement of a module due to electrical failure, you must
perform the adjustment procedure if the module you replaced is marked “yes” in
the following table. (See Table 3-1.)
Table 3-1: Adjustments required for module replaced
Module replaced
Adjustment required
Hard disk assembly
No
Fan tray unit
No
Front panel assembly
No
Drive bay module
No
MIO board
Yes
Display Adapter board
No
Processor module
No
Power supply
No
Power board
No
CLK12G module
Yes
AWG12G board
Yes
OUT10G board
Yes
AWG7000B and AWG7000C Series Service Manual
3–1
Summary of Contents for AWG7 1B Series
Page 2: ......
Page 6: ......
Page 16: ...Environmental Considerations x AWG7000B and AWG7000C Series Service Manual...
Page 18: ...Preface xii AWG7000B and AWG7000C Series Service Manual...
Page 19: ...Operating Information...
Page 20: ......
Page 22: ...Operating Information 1 2 AWG7000B and AWG7000C Series Service Manual...
Page 23: ...Theory of Operation...
Page 24: ......
Page 33: ...Adjustment Procedures...
Page 34: ......
Page 52: ...Adjustment Procedures 3 18 AWG7000B and AWG7000C Series Service Manual...
Page 53: ...Maintenance...
Page 54: ......
Page 130: ...Troubleshooting 4 76 AWG7000B and AWG7000C Series Service Manual...
Page 131: ...Replaceable Parts...
Page 132: ......
Page 136: ...Replaceable Parts Figure 5 1 Cosmetics 5 4 AWG7000B and AWG7000C Series Service Manual...
Page 166: ...Replaceable Parts Figure 5 18 CLK12G module 5 34 AWG7000B and AWG7000C Series Service Manual...
Page 168: ...Replaceable Parts Figure 5 19 AWG12G module 5 36 AWG7000B and AWG7000C Series Service Manual...
Page 170: ...Replaceable Parts Figure 5 20 SMA bracket 5 38 AWG7000B and AWG7000C Series Service Manual...
Page 172: ...Replaceable Parts Figure 5 21 Relay unit 5 40 AWG7000B and AWG7000C Series Service Manual...
Page 174: ...Replaceable Parts Figure 5 22 Rear connectors 5 42 AWG7000B and AWG7000C Series Service Manual...