Install the AWG-HD into a system
e.
Send an *IDN? query to ensure the device responds. If communication is
established correctly, the device should respond with its information in
the Command/Data
fi
eld.
If the instrument does not appear in the Instruments, it may need to cycle
power one or twice before its host name appears. Allow at least 30
seconds after power on for the instrument to be identi
fi
able.
If you have problems discovering the instrument with OpenChoice, you
can use the TekExpress HDMI option to identify the instrument.
NOTE.
The AWG-HD has a fallback IP address in case normal connection
through its host name fail. If the AWG-HD is connected directly to a host PC
via Ethernet (Link-Local) without the use of a network using DHCP, it will use
an IP address of 192.168.1.10.
6.
Set up all of the instruments using Instrument Control Settings in TekExpress.
This completes the hardware setup for the HDMI test system.
7.
See the TekExpress Help to complete the software setup.
AWG-HD Installation and Safety Instructions
7
Summary of Contents for AWG-HD
Page 1: ...xx AWG HD Installation and Safety ZZZ Instructions P071365200 071 3652 00 ...
Page 2: ......
Page 3: ...AWG HD Installation and Safety ZZZ Instructions xx www tek com 071 3652 00 ...
Page 25: ...시스템에 AWG HD 설치 그림 3 교정 설정 그림 4 DUT 설정 AWG HD Installation and Safety Instructions 19 ...
Page 40: ...将 AWG HD 安装到系统 图 5 校准设置 图 6 DUT 设置 34 AWG HD Installation and Safety Instructions ...
Page 55: ...將 AWG HD 安裝至系統中 圖表 7 校驗設定 圖表 8 DUT 設定 AWG HD Installation and Safety Instructions 49 ...
Page 70: ...AWG HD型のシステムへの組込み 図 9 校正セットアップ 図 10 DUTセットアップ 64 AWG HD Installation and Safety Instructions ...