AFG1000 Series Arbitrary Function Generator Programmer Manual
Command Descriptions
Technical Reference
.
Conditions
For the AFG1022, use
[SOURce1]
. Modulation, sweep, and burst conditions do
not apply to CH2 for the AFG1022.
Group
Source
Syntax
[SOURce[1|2]]:FREQuency:STOP
{<frequency>|MINimum|MAXimum}
[SOURce[1|2]]:FREQuency:STOP
? {MINimum|MAXimum}
Related Commands
Arguments
<frequency>::=<NRf>[<units>]
where:
<NRf> is the stop frequency.
<units>::=[Hz | kHz | MHz]
Returns
<frequency>
Examples
SOURce1:FREQuency:STOP 100KHz
sets the stop frequency of CH 1 to 100 kHz.
[SOURce[1|2]]:FSKey[:FREQuency]
This command sets or queries the hop frequency of FSK modulation for the
specified channel.
Conditions
For the AFG1022, use
[SOURce1]
. Modulation, sweep, and burst conditions do
not apply to CH2 for the AFG1022.
Group
Source
Syntax
[SOURce[1|2]]:FSKey[:FREQuency]
{<frequency>|MINimum|MAXimum}
[SOURce[1|2]]:FSKey[:FREQuency]
? {MINimum|MAXimum}