Operating Basics
Overview
To perform optical compensations
Control elements and resources
Run the user
wavelength
gain
compensation
You can optionally use a custom input signal to compensate
an optical channel:
NOTE.
You must know the optical power value of the
custom signal. Use an independently calibrated average
optical power meter to precisely measure and record the
custom optical signal power. Then connect the signal to the
module using the same
fi
ber cables.
6.
In the Vert Setup dialog box, click the
User
Wavelength Gain
button under Compensation. Follow
the on-screen instructions.
Set the wavelength and power values of the signal
to be applied to the channel in the User Wavelength
Gain Compensation dialog box.
7.
Click
OK
to execute the compensation.
8.
Repeat steps 2, 6, and 7 to compensate additional
optical channels.
80C12B Optical Sampling Module User Manual
17