Getting Started
5.
Turn the hold-down screws clockwise to lock the module in place.
6.
Once you have installed the module, power on the instrument. Verify that
the module passes power-on tests.
NOTE.
When
fi
rst installing a sampling module(s) or after moving a sampling
module from one compartment to another, run a module compensation (Utilities
> Compensation) to ensure that the instrument meets it speci
fi
cations. You must
also run a compensation if an extender is installed, changed, or removed from a
module. (See page 14, Optimizing Measurement Accuracy.)
After running compensation, save the new values to retain them; otherwise they
are lost when powering off the instrument.
Removing a Module
1.
Power off the instrument using the front-panel On/Standby power switch.
2.
Plug the grounding strap into the instrument ground connector, and place the
ground strap on your wrist, with contact to skin.
3.
Turn the hold-down screws all the way counterclockwise so that they are
completely out and the module retaining tabs are
fl
ush with the edge of the
module.
4.
Slide the appropriate large module ejector lever sideways to unseat the
module from the mainframe connector.
5.
Pull on the hold-down screws to remove the module from the slot.
6.
Handle the module appropriately. For example, move it to another slot in the
instrument or place it in a static-protected environment for transport or storage.
8
80C12B Optical Sampling Module User Manual