fi
cation
Setup
Figure 14: DC calibration output test setup
1.
Install an 80E0X module. Power on the instrument and wait 20 minutes for
warm-up.
2.
Connect the test equipment (See Figure 14.)
3.
Push the front-panel
DEFAULT SETUP
button and click
Yes
in the
con
fi
rmation dialog box.
4.
Select
Setup > Mode/Trigger
and click
Direct
trigger mode.
5.
Set the multimeter to read
DC volts
.
Test
Follow this procedure to make the DC calibration output check:
1.
Select
Setup >Vertical
.
2.
Set the
DC Cal
parameter to each of the following settings and record the
DMM readings in the test record.
+1.0 V
0.0 V
–1.0 V
3.
Check the readings against the limits in the test records.
4.
Disconnect the test setup.
End of Procedure