![Tektronix 80A00 Technical Reference Download Page 180](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946180.webp)
DSA8300 Performance Veri
fi
cation
Setup
Figure 10: Phase corrected untriggered jitter test setup
1.
Install the 82A04 Phase Reference module into the right-most small module
slot of the instrument.
2.
Install the 80E0X module(s) to test into the left-most small module
compartments of the instrument as shown in the
fi
gure. Power on the
instrument and wait 20 minutes for warm-up.
3.
Connect the test equipment (See Figure 10.)
4.
Set the synthesized signal generator as follows:
Frequency to
10.0 GHz
.
Output to
1.6 V
p-p
(800 mV
p-p
displayed).
5.
Set the DSA8300 as follows:
a.
Push the
Default Setup
front-panel button and click
Yes
.
b.
Select the input channel of the module under test from the Waveform
Selector menu (left end of controls bar at bottom of the graticule).
c.
Select
Setup > Mode/Trigger
and set the Trigger Source to
Free Run/
Untriggered Phase Ref
.
d.
Select
Setup > Phase Ref
and set the following:
Phase correction mode to
Untriggered
.
Source to
C7_C8
.
Input freq to
10.00 GHz
.
Click the
Characterize
button and follow the on screen instructions.