Calibration
—
7623/R7623
Service
HORIZONTAL
DEFLECTION
SYSTEM
Equipment
Required
1. 7B53A plug-in unit
2. 7A15A plug-in unit (two)
3.
Mainframe standardizer
calibration fixture
4. Test-oscilloscope
system with two 10X probes
5.
Time-mark generator
6.
Low-frequency generator
7. Medium-frequency generator
8.
Dual-input
coupler
9.
Five-nanosecond G R cable
10. 50-ohm GR
in-line termination
11. 42-inch 50-ohm
BNC cable
12. 50-ohm BNC
termination
13.
Three-inch
screwdriver
14. Low-capacitance screwdriver
Control
Settings
Set
the controls
as
given in the Preliminary Control
Settings.
22.
Adjust
Horizontal
Amplifier
Limit
Centering
a.
Remove
the
mainframe standardizer calibration fix
ture
and 7B53A. Install the
7B53A in the horizontal
compartment.
b.
Set
the 7B53A for auto, internal triggering at a sweep
rate of one millisecond/division with the magnifier
on.
c. Connect
10X probes to
both inputs of the test
oscilloscope.
Connect the probe
tips to the horizontal
deflection
plate
connectors of
the 7623 (be sure probes are
compensated).
d. Set
both channels
of the test oscilloscope for a
vertical
deflection factor of
0.5
volt/division (five volts/
division
at
probe tip) in the chop dual-trace mode with the
input
coupling set to ground.
e.
Position
the ground-reference traces displayed on the
test
oscilloscope to the center horizontal
line of the
graticule.
Do not
change the
test-oscilloscope position
controls after establishing this
ground reference.
f. Set the test oscilloscope for DC
input coupling and
set
the triggering
controls so that the test oscilloscope is
triggered
from the
signal on
channel 1 only. Set the
triggering
controls for a
stable
display at
a sweep rate
of
two
milliseconds/division.
g.
CHECK-The
baseline of both displayed waveforms
should be
at
the same DC level within 0.2 division (see Fig.
2-5).
h.
ADJUST
—Limit
Centering adjustment
R535
(see Fig.
2-6)
to match the DC levels of both waveforms.
i.
INTERACTION-lf
R535 is adjusted,
re-check
steps
22
through 25.
j.
Disconnect all
test equipment.
23.
Adjust
Horizontal
Amplifier
Centering
a.
Set
the Test
switch on the calibration
fixture
to
TRIGGERING
GAIN.
2-17
Summary of Contents for 7623
Page 1: ...MANUAL 7623 R7623 STORAGE OSCILLOSCOPE SERVICE MANUFACTURERS OF CATHODE RAY OSCILLOSCOPES ...
Page 51: ...Fig 3 2 Block diagram of Logic circuit Circuit Description 7623 R 7623 Service ...
Page 72: ...W NJ 00 Fifl 3 22 Low Voltage Power Supply detailed block diagram ...
Page 73: ...Circuit Description 7623 R 7623 Service ...
Page 74: ...CO NJ CD Fig 3 22 Low Voltage Power Supply detailed block diagram cont ...
Page 75: ...Circuit Description 7623 R 7623 Service ...
Page 97: ...Circuit Description 7623 R7623 Service 3 51 ...
Page 98: ...Circuit Description 7623 R7623 Service Fig 3 39 Output Pulses for the Storage Circuits 3 52 ...
Page 103: ...Circuit Description 7623 R7623 Service 3 57 ...
Page 108: ... Ç À Fig 4 2 Location of circuit boards in the 7623 ...
Page 109: ...Fig 4 3 Location of circuit boards in the R7623 Maintenance 7623 R 7623 Service ...
Page 113: ...Maintenance 7623 R7623 Service Fig 4 6 Circuit Isolation Troubleshooting Chart 4 9 ...
Page 165: ...7623 BLOCK DIAGRAM ...
Page 166: ...7623 R7623 Service Front of Board ...
Page 167: ......
Page 168: ...FL ...
Page 169: ......
Page 173: ...7623 Logic ...
Page 175: ...Vertical Interface A4 ...
Page 178: ...Vertical Interface ...
Page 180: ...Vertical Amp A5 ...
Page 184: ...Horizontal Amp A6 ...
Page 186: ...7623 TO P450 VERT AMP 3 HORIZONTAL AMPLIFIER ...
Page 188: ...Output Signals A7 ...
Page 190: ...FROM 7G23 Output Signals g ...
Page 195: ...FROM LV POWER SUPPLY 7623 CRT CIRCUIT ...
Page 197: ......
Page 200: ......
Page 202: ...Storage Output A14 ...
Page 204: ......
Page 205: ...7623 R7623 Service Fig 6 14 A15 Cal Storage circuit board ...
Page 208: ......
Page 209: ...7623 R7623 Service Fig 6 15 A16 Readout System circuit board ...
Page 213: ......
Page 242: ...7623 R7623 OSCILLOSCOPE b ...
Page 243: ...151 ...
Page 246: ...20 ...
Page 247: ...7623 R7623 OSCILLOSCOPE 112 ...