Section 2—
634 Operators
OPERATING
INSTRUCTIONS
AMBIENT TEMPERATURE
CONSIDERATIONS
This instrument
can be operated
where the ambient air
temperature
is between 0° and +50° C (+32° and
+122°
F), and
can be
stored in
ambient temperatures between
-40°
and
+70° C (-40° and +1 58° F). After being
stored in
temperatures
beyond the above operating limits,
allow
the
chassis temperature to return to within the operating
limits
before
applying
power. Allowing the Monitor to
operate at
an ambient
temperature substantially higher
than
that
specified may result in poor reliability as well
as inaccurate performance.
When
the 634
is mounted
in a rack
with other
equipment,
it is important that the ambient temperature
surrounding the
Monitor does not 50° C (+122°
F). Additional clearance for convection,
or forced
ventilation methods
(fan), may be needed to maintain
ambient
temperatures below +50° C (+122° F). Reliability
and
performance of the 634 will be affected if the
ventilation holes
in the protective panels are
obstructed,
or
if the
634 is operated at an ambient temperature
higher than
+50° C (+122°
F). Other environments and
mounting
configurations may require additional cooling
measures.
CONTROLS AND CONNECTORS
Controls and
connectors necessary for operation of the
634 Monitor
are located on the front
and
rear panels of
the
instrument.
To make full use of the capabilities of this
instrument,
the
operator
should be familiar with the
function
and use of each external control and connector.
The
front-panel
controls are shown and described in
Figure
2-1. Brief descriptions of the standard rear-panel
controls
and connectors are given in Figure 2-2. The
Option
20 rear panel
is illustrated
in Figure 2-3.
INPUT
SIGNAL REQUIREMENTS
The
634 Monitor requires a negative sync-tip video input
signal
for proper operation. The input signal amplitude
range should not exceed
0.35 to 2 volts peak-to-peak for
normal operation,
or a
maximum of
5
volts peak.
OPERATORS
CHECKOUT
PROCEDURE
The
following
procedure is provided to aid in obtaining a
display on
the 634 Monitor and may be used as
a check
of
basic
instrument operation. The procedure may be
used for incoming
inspection to verify
proper operation,
and
may also
be used
by
the operator for instrument
familiarization.
Only
instrument functions, and not
measurement
quantities or specifications,
are checked in
this
procedure. Therefore, a minimum amount of
test
equipment
is required. If performing the Operators
Checkout
Procedure reveals improper performance or
instrument
malfunction, first check the operation of
associated
equipment, then refer to qualified service
personnel
for
repair or adjustment of the instrument.
TEST EQUIPMENT
REQUIRED
The
following
test equipment
was used as a basis to
write
the Operators Checkout Procedure. Other test
equipment,
which
meets these requirements, may be
substituted. When
other
equipment is substituted, the
control
settings
or
set up may
need to be altered.
1. Sync
Pulse and Test Signal
Generator
Description:
Negative-synchronized
video signal
generator.
Type
Used:
TEKTRONIX
1470 NTSC Color Sync Test
Signal Generator.
2.
Cable
Description:
Coaxial;
length,
42 inches; connectors,
BNC
male-to-male; impedance, 75 ohms.
Type
Used: Tektronix
Part 012-0074-00.
PRELIMINARY SETUP
1. Connect
the 634 Monitor power cord to a suitable
power
source.
NOTE
Check
the
line
voltage
information indicated
on
the rear panel.
If the
source voltage is not
within this range, refer
qualified service
personnel
to the servicing
information
sections
of
the 634 Instruction Manual.
2. Open the access
door on the front panel
and set the
controls
as
follows:
BRIGHTNESS
......... .................
Fully
counterclockwise
FOCUS
.................... Midrange
CONTRAST
..........
Midrange
POWER
............................. Out (PULL ON)
3.
Allow
at least
one
minute
for
the
instrument
to
warm
up.
2~1