SECTION
II
OPERATING
INSTRUCTIONS
The Type
511-A may be
operated
in
any
normal
in
door
location,
or
in
the open
if
protected
from
moisture.
If
the
instrument
has
been
exposed
to dampness,
it
should
be
left
in
a
warm
room until
thoroughly dry,
be
fore
being
placed in
operation.
To
prevent
excessive
interior temperatures,
it
is
im
portant
to
allow
adequate ventilation
for
the
oscillo
scope.
Several
inches
clearance
should
be
provided
at
the
sides, back
and
top
of
the
case.
If
the
Type
511-A
is
used
continuously
for
the
same
application,
and
only
one
setting
of
the
controls
is
re
quired,
it
is
advisable
to
periodically check the
instru
ment
at
all
control
settings
to
be
sure
it is
in
normal
operating
condition.
For
example,
if
only
direct
connec
tion
to the
vertical
deflection
plates
of
the
CRT is
used,
a
defect
in the
video amplifier
might
not
become
evi
dent.
Operation
of the
controls also
helps
to
prevent
accumulation
of
dirt
and
tarnish
on
their
contacts.
Although
the
components
are
well
supported
and
the
adjustments
very stable
in
setting
to allow
for
portable
operation,
the
Type
511-A
should
not
be subjected to
excessive
vibration
or
rough
handling.
To
place
the
Type
51
1-A
in
operation
for the
first
time
the
following
procedure
is
suggested.
1.
Connect
to
a
source
of
50-60
cycle,
105-125V
power.
2.
Set
controls
as
indicated
below.
3.
Turn
POWER switch
to
ON
and
wait
about
30
seconds.
INPUT
CHAN.
SEL............. ..................... 1 STAGE
INPUT
ATTENUATOR
... ................................
X1
VERT.
AMPL.
ATTEN
....... ...................Clockwise
SIGNAL
CALIBRATE .... ................................
X
1
VERT.
POSITION..
........ .............
Index
vertical
CAL. VOLTAGE
.............. ....................................
0
ASTIGMATISM
.................. ............ Index
vertical
INTENSITY
...................... .... Counterclockwise
FOCUS
................................ ..............Index vertical
HOR.
POSITION
..................................... Clockwise
SWEEP
MAGNIFIER
POSITION
...................
OUT
SWEEP
SPEED
MULT....
.....................................
5
SWEEP
RANGE................ ..........................
10,000
SWEEP
STABILITY
.......... ....
Counterclockwise
TRIGGER
AMPL...............
...
.Counterclockwise
TRIGGER
SEL
................... ...........................
+INT.
4.
Advance
INTENSITY
control until
a
spot
is seen.
5.
Adjust
VERT.
POSITION,
HOR.
POSITION, ASTIG
MATISM
and
FOCUS
until
a
sharply
focussed
spot
is
obtained at
the
left
center
of the
screen.
CAUTION
—
DO
NOT
ALLOW
THIS
SPOT
TO
BE
EXCESSIVELY
BRIGHT
OR
RE
MAIN FOR
LONG
IN
ONE
POSITION.
6.
Advance
the
SWEEP
STABILITY
control
until
a
sweep
appears,
then
turn
it
back
just
under
this
point.
7.
Set
the
CAL.
VOLTAGE
control
to
10.
A
vertical
line
about
1
½
inches
high
should
appear.
8.
Advance
the
TRIGGER
AMPL.
control
until
a
stable
image
of
a
sine
wave
appears.
The
oscilloscope is
now
observing
the
50-60
cycle
power
line
waveshape.
To
observe
other
waveshapes
connect
them
to
the
SIGNAL
INPUT
binding
post,
turn
the
SIGNAL
CALIBRATE
switch
to
SIGNAL,
select
the
appropriate
sweep
speed,
etc.
SWEEP
CIRCUIT
ADJUSTMENTS
The
sweep
circuit
of
the
Type
511
A
is
different
from
the
gas tube
type
usually employed
in
portable oscillo
scopes.
It
is
much
more
flexible
and
capable of syn
chronizing
with much
higher
frequencies than
conven
tional
sweep
circuits.
Fundamental frequencies of
10
mc
and
higher can be
easily
observed.
By
one
simple
adjustment
the
sweep
will
run
recurrently
or
triggered
as
desired.
In
a
recurrent
or
sawtooth
sweep
as
generated
by
a
gas
tube
or
similar device
the
synchronization
takes
place
at
the
end
of
the
sweep.
The sync
impulse
causes
the
beam to return
to
the
left
side
of
the
tube
where it
immediately
starts
another
sweep.
Thus
the
sweep
time
must
be
a
multiple
of
the
signal
period,
or
expressed
another
way,
the
sweep
frequency
an
integral
fraction
of
the
signal
frequency.
In
a
triggered
sweep circuit
each
sweep
is
started
in
dependently
of
the
preceeding
sweep
by
a trigger
or
synchronizing impulse.
When
no
trigger
is
being
re
ceived,
the
beam
remains
at
the
left
side
of
the
trace.
When
the
trigger
arrives, the
beam
goes
linearity
to
the
right
for
a
time
in
this
case
determined
by
the
multi
vibrator
pulse
length.
At
the
end
of its sweep
it
returns
to the
left
side again to
await
another
trigger.
It
is
this
variable
waiting
period
which
makes
the
synchroniza
tion
so
easy
since
the sweep time
is
now
independent
of
the
signal
period.
The
triggered
sweep
makes
it
pos
sible
to
view pulses such
as
used
in
radar and tele
vision
where the length
of
the
pulse
is
very
short
com
pared
to
the
space
between pulses.
This
requires
a
fast
sweep
with
a
long
waiting period.
Waiting periods
of
2000
times
the
sweep length are frequently used.
ADJUSTMENT
OF
SWEEP
STABILITY
The
SWEEP STABILITY control
varies
the
multivibra
tor
bias
and
therefore
detemines whether the sweep
will
oscillate
recurrently
or
wait
for
a
trigger
impulse.
Correct
functioning
for
triggered operation is
therefore
obtained
when
it
is
set
just
under
the
point
where oscil
lation
takes
place.
The
procedure
is
to remove the
trigger
source,
turn
the
TRIGGER
AMPL.
control
counter
clockwise,
and
advance
the
SWEEP
STABILITY
knob
until
a sweep appears and then
decrease
it
until
it
just
disappears.
SECTION
2, PAGE
1
TYPE
511-A,
511-AD
Summary of Contents for 511A
Page 1: ......
Page 2: ......
Page 6: ......
Page 10: ......
Page 36: ......
Page 40: ...Fig 42 Interior view right front side SECTION 4 PAGE 22 TYPE 511 A 511 AD ...
Page 41: ...X Fig 43 Interior view left side SECTION 4 PAGE 23 TYPE 511 A 511 AD ...
Page 42: ...E CI5E Fig 44 Sweep chassis bottom view R63 C2I TYPE 511 A 511 AD SECTION 4 PAGE 24 ...
Page 43: ...SECTION 4 PAGE 25 TYPE 511 A 511 AD ...
Page 47: ...Km SECTION 4 PAGE 29 TYPE 511 A 511 AD C9 CIOA v LI 1 L LP4 ...
Page 48: ......
Page 56: ......
Page 58: ......
Page 59: ...TRIGGER AMPLIFIER MULTIVIBRATOR SWEEP GENERATOR SWEEP MAGNIFIER SWEEP AMPLIFIER ...
Page 60: ......
Page 62: ......
Page 64: ......
Page 68: ......
Page 70: ......
Page 71: ......