
NOTE
—
When
using
the
internal
trigger im-
pulses
to view fundamental
frequencies
above
about
2mc
it
will
be
necessary
to
progressively
advance
the
SWEEP
STA
BILITY control
somewhat
above
the
normal
operating
point
to
secure
a
stable
pattern.
Function
of
Trigger
Sei.
This
control
selects
the
trigger
impulse
to
be
used
from
three
sources,
the
line
frequency,
(60
cycles),
the
video
amplifier
(
+INT
and
—
INT),
or
the
TRIGGER IN
PUT
binding
post.
Two
positions
are available
for
both
the
INT. and EXT.
sources. When using
those
marked
4-
the
sweep
starts
at
the
rising
portion
of
the
trigger
impulse.
The
—
positions
start
the sweep
on
the
falling
portion.
When
the
INPUT
CHAN.
SEL.
switch
is
in
the
2
STAGES
position
the +
INT
and
—
INT
positions
are
reversed
in
polarity
since
the
trigger impulses
are
shifted
180
º
by
the
pre-amplifier
stage.
For
satisfactory
opera
tion
in
the
EXT.
positions
a
trigger
of
from
1
V
to
50
V
peaks
should be
provided.
Larger
triggers
should
be
reduced
by
an
external
attenuator.
Adjustment
of
Trigger
Ampl.
The
TRIGGER
AMPL. control
adjusts
the
bias
in
the
trigger
amplifier
and
therefore
the
amplitude of
the
impulse
which
reaches
the
multivibrator.
After
the
SWEEP STABILITY
is
set
as
described
previously,
and
the
TRIGGER
SEL.
switch
set
to the
correct
position,
ad
vance
the
TRIGGER AMPL. control
until
a
stable
image
is
obtained.
When
the
trigger
source
is
a
slowly
chang
ing
wave
such
as
a
sine
wave
or
sawtooth, adjustment
of
the
TRIGGER
AMPL.
setting
will
change
the
phase
of
the
signal
relative
to
the
start
of
the
sweep.
In
general,
it
is desirable to
use
the
minimum
TRIG
GER AMPL. setting
required to
insure
a
stable
image.
If
poor
synchronizing
is
experienced
when
the
Type
511
-A
is
used
to
observe
repetitive signals containing
erratic
peak
voltages,
such
as
produced
by vibrating
contacts,
it may
be
possible
to
form
a
stabilized
trigger
signal
by
the
use of
auxiliary
limiting
and
differentiat
ing
circuits.
The
TRIGGER
SEL.
switch
should then
be
set
to
the
proper
EXT.
polarity,
and
connection
made
to
the
TRIGGER
INPUT.
Adjustment
of
Sweep
Speed
The
combination
of
SWEEP
RANGE and
SWEEP
SPEED MULT,
controls
allows
the
operator
to
vary
the
sweep speed
so
that
the
marked
portion
of the grati
cule
(10
cm)
is
covered by
the
trace
in
any
time
from
.1
sec. to
1
microsecond.
The
number
of
microseconds
required
for
the
sweep
to
traverse
the
10
cm.
scale
is
determined
by multiplying the SWEEP
RANGE reading
by
the
sum
of
the
readings
of
the
SWEEP SPEED MULT,
dials.
This
indicated
reading
is
accurate
within
5%
when calibrated
at
the
factory.
Provision
has been
made to
compensate
for
variations
caused
by different
tubes,
etc.,
by
a
screwdriver
adjustment
inside
the
case.
Procedure
for adjustment
of
this
control
is
described
in
Section IV.
Recurrent Sweep
In
case
it
is
desirable
to
have
a
sweep
without
using
any
sort
of
trigger,
merely
advance
the
SWEEP
STABILI
TY
control
until
a
stable
sweep
is
obtained.
This
sweep
is
the
conventional
sawtooth
variety
as
used
in
most
oscilloscopes.
Single
Sweep
The
triggered sweep
circuit
used
in
the
Type
511
in
herently
provides
for
single
sweep
operation.
The
beam
is
blanked
out
until
the
trigger simultaneously
turns it
on and
starts the sweep.
For
single
sweep
operation
a
mechanical
contact
or
hand
key
can
be
connected
be
tween
the
TRIGGER
INPUT
binding
post
and
ground.
If
repeated sweeps
occur
due
to intermittent
or
bouncing
contact, a
capacitor
of .1
mf.
to
1
mf. should
be
con
nected across
the
contact.
This
capacitor
is
charged
through
22
megohms
and
the
long
time constant
pre
vents
an
immediate
recurrence of
the
sweep.
Sweep
Magnifier
Frequently
it
is
desirable
to
examine
in
some
detail
a
portion
of
the
waveshape
under observation,
for
in
stance the
rate
of
fall
of the
end
of
a
fairly
long
pulse.
In
the
Type
511-A
the
sweep magnifier circuit
makes
it
possible
to
expand
any
desired
20%
of
the
sweep
to
cover
the entire
tube
face.
When
the
SWEEP
MAGNI
FIER
POSITION
control
is
moved
from
the
OUT
position,
the
sweep
is
delayed
for a variable
time
and
then
goes
at
about
5
times
its
normal rate.
The
operating
pro
cedure
is
to
turn the
SWEEP
MAGNIFIER
POSITION
to
OUT.
Adjust
the
sweep
speed
controls
so
that
the
por
tion
of the
wave
to
be
magnified is
at
the
right
side
of
the
tube.
Now
turn
the
SWEEP
MAGNIFIER
POSITION
knob
clockwise
until
the
desired
signal moves in
from
the
right to
the
center of
the
tube
face.
The
magnified
sweep
is
not
linear over
its
entire
trace
so
that
it
is
de
sirable
to
use only
the
center
two
inches
if
possible.
If
higher
magnification
than five is
wanted
it
is
only
neces
sary
to
change
R47
to
a
larger
value.
A
value
of
10
K
will
give
a
magnification
of
about 10.
The
somewhat
lower
value
chosen
appears to
be
a
more
usable
value
in
most cases.
NOTE
—A
sweep
of
10
cm
per
microsecond
is
about
the
limit
of
the
sweep
amplifiers,
so
that
care
must
be
taken
that
the
magni
fied
sweep
does
not
exceed
this
value.
Ext.
Sweep
Input
Provision
has
been
made
for
the
connection of
ex
ternal
sweep
generators
to
the
Type
511A.
This
might
be
a
sine
wave
oscillator
for
frequency
comparison,
the
sweep
obtained
from
a
"Frequency
sweep"
signal
gen
erator, etc.,
or an
auxiliary
slow
sweep
generator. The
EXT.
SWEEP
INPUT
binding
post is
connected
to
the
horizontal
deflection
plates
via
the
EXT.
SWEEP ATTEN.
and
a
one
stage
push
pull
amplifier.
The
entire
system
is
DC
coupled,
permitting
the
use
of
very
slow
sweeps.
TYPE
511-A, 511-AD
SECTION
2,
PAGE 2
Summary of Contents for 511A-AD
Page 1: ......
Page 2: ......
Page 4: ......
Page 8: ......
Page 13: ...2 7 0 180 120 6 0 DEGREES OF PHASE SHIFT FREQUENCY TEKTRONIX JNC 12 28 50 ...
Page 14: ......
Page 20: ......
Page 41: ...Fig 41 Input Probe disassembled SECTION 4 PAGE 21 TYPE 511 A 511 AD ...
Page 42: ...Fig 42 Interior view right front side TYPE 511 A 511 AD SECTION 4 PAGE 22 ...
Page 43: ...Fig 43 Interior view left side SECTION 4 PAGE 23 TYPE 511 A 511 AD ...
Page 44: ...TYPE 511 A 511 AD SECTION 4 PAGE 24 ...
Page 45: ...Tf S 4 t C15C C15D ...
Page 46: ...R30 1 SECTION 4 PAGE 25 TYPE 511 A 511 AD ...
Page 48: ...SECTION 4 PAGE 27 TYPE 511 A 511 AD ...
Page 50: ...SECTION 4 PAGE 29 TYPE 511 A 511 AD L24 ...
Page 51: ......
Page 61: ......
Page 62: ...EXT BLOCK DIAGRAM OF TYPE 5IIA CATHODE RAY OSCILLOSCOPE 7 13 51 ...
Page 63: ......
Page 64: ...SWEEP MAGNIFIER SWEEP AMPLIFIER ...
Page 65: ......
Page 67: ......
Page 69: ......
Page 73: ......
Page 74: ......
Page 75: ......