Malntenance-5110
Multimeter
Description:
Voltmeter, 10
megohm input impedance
and
0 to 250 volts range; accuracy, within
0.1%.
Ohmmeter,
0 to 20 megohms; accuracy, within 3%. Test
probes must
be insulated to prevent accidental shorting.
Purpose:
To check voltages and for general
troubleshooting in
this
instrument.
Recommended
Type;
TEKTRONIX DM 501
Digital
Multimeter (requires a TM 500-series power module).
Test
Oscilloscope
Description:
Frequency
response,
de to 2 megahertz
minimum;
deflection
factor, 1 millivolt to 5 volts/division. A
10X, 10 megohm voltage probe
should be used to reduce
circuit loading.
Purpose:
To check operating waveforms and for
general
troubleshooting.
Recommended
Type:
TEKTRONIX 5110, 5A13N,
5B10N oscilloscope system
or equivalent.
Use a
TEKTRONIX P6108
or P6062A Probe.
3.
Visual
Check.
Visually check the portion of the
instrument in which the
trouble is located. Many troubles
can
be located by visible indications such as unsoldered
connections,
broken wires,
damaged circuit boards,
damaged
components, etc.
4.
Isolate
Trouble
to
a Circuit.
To
isolate trouble to a
circuit,
note the trouble symptom. The symptom often
identifies the circuit in which the trouble is located. When
trouble
symptoms
appear in more than one circuit, check
the affected circuits by
taking voltage and waveform
readings. Incorrect
operation of
all circuits often indicates
trouble in the
power supply. Check first for correct
voltages
of the individual supplies. See Table 4-3.
However,
a defective component elsewhere in the instru
ment can
appear as a power-supply
trouble and may also
affect
the operation of other circuits.
Table
4-3
POWER SUPPLY OUTPUT VOLTAGES
Power
Supply
Output
Voltage
Range
Typical
120 Hz
Ripple
+200 V
+175
V to
+247.5 V
3 V or less
+30 V
+29.82 V
to+30.18 V
3 mV or
less
+5 V
+4.89 V
to +5.11
V
2 mV
or less
-30
V
-29.89
V to -30.11
V
2 mV
or
less
TROUBLESHOOTING
TECHNIQUES
The following troubleshooting procedure is arranged
to
check the simple trouble possibilities before
proceeding with
extensive troubleshooting.
The first few
checks ensure
proper connection,
operation, and adjust
ment.
If the trouble
is not located by these checks,
the
remaining
steps aid in locating the
defective component.
When
the defective component is located, it should be
replaced
using the replacement procedure given under
Corrective
Maintenance.
5.
Check
Voltages and
Waveforms.
Often
the
defective
component
can
be located by checking for the correct
voltages and waveforms in
the circuit.
6.
Check Instrument
Adjustment.
Check
the adjust
ment
of this instrument, or the affected circuit if the
trouble appears in one circuit. The apparent
trouble
may
be
the result of misadjustment. Complete adjustment
instructions are given in
Section 3.
Troubleshooting
Procedure
1.
Check
Control
Settings:
Incorrect
control
settings
can
indicate
a
trouble that does not exist.
If there is any
question
about the correct function or operation of any
control,
see Fig. 1-2 in the Operating Instructions section.
2.
Check
Associated Equipment, Before trouble
shooting, check that the equipment used with
this instru
ment is properly connected and that the interconnecting
cables
are not defective.
Also,
check the power source.
7.
Check Individual Components.
The following
procedures
describe
methods for checking individual
components.
Two-lead components that are soldered In
place
are best checked
by first disconnecting one end.
This
isolates the measurement from the effects of surroun
ding circuitry.
To
avoid
component damage, disconnect the power
source
before
removing or replacing semiconduc
tors.
4-10
REV
A, NOV
1978
Summary of Contents for 5110
Page 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Page 22: ......
Page 34: ......
Page 42: ......
Page 83: ...5110 I43XII3 2134 15 ...
Page 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Page 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Page 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Page 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Page 95: ...DEFLECTION AMPLIFIERs ...
Page 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Page 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Page 114: ......
Page 115: ...5110 OSCILLOSCOPE ...
Page 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Page 117: ...1 ...
Page 121: ...o ...
Page 122: ...S 1 5110 OSCILLOSCOPE ...
Page 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Page 124: ......