Operating Instructions— Type 503
TABLE 2-1
SLOPE
Determines whether
triggering occurs on
rising portion ( + slope) or falling portion
(— slope) of triggering waveform.
COUPLING
Selects whether triggering w ill occur at
a specific DC level or a t an AC level.
SOURCE
Selects the source of the triggering signal.
EXTERNAL TRIG. Input connector for external triggering
IN
signals.
DC BAL.
+INPU T and
-IN P U T con
nectors
+INPU T and
— INPUT
switches
Provides for vertical stability o f no-signal
trace for all positions o f the SENSITIVITY
control.
Input connectors for signal to be displayed
vertically. (Signals connected to the — IN
PUT connector w ill be displayed inverted.)
Select desired coupling for incoming sig
nal.
HORIZONTAL CHANNEL
LEVEL
Selects the voltage point on the trigger
ing waveform at which the horizontal
sweep is triggered. This control also selects
automatic triggering (AUTO, position) or
allows the sweep to free run (FREE RUN
position.)
SWEEP TIME/CM Selects the desired horizontal sweep rate
from a choice of 21 calibrated steps.
SWEEP TIME/CM Provides a continuous range of sweep
VARIABLE (red)
rates between the fixed steps selected by
the SWEEP TIME/CM switch. (The sweep
rates are calibrated only when the VARI
ABLE control is set fully clockwise to the
CALIBRATED position.)
SWEEP STABIL-
Provides for a stable presentation when
ITY ADJUST
the LEVEL control is in the AUTO, posi
tion.
DC BAL.
Provides for horizontal stability of no
signal trace for all positions of SENSI
TIVITY control.
SENSITIVITY
Selects the sensitivity of the Horizontal
Am plifier from 14 calibrated steps.
SENSITIVITY
Provides a continuous range o f sensitivity
VARIABLE (red)
values between the fixed calibrated steps
selected by the SENSITIVITY control. (Nor
mally, sensitivity is calibrated only when
VARIABLE control is in CALIBRATED posi
tion.)
POSITION
+INPU T and
-IN P U T con
nectors
Controls horizontal positioning of the dis
play on the screen.
Input connectors for signal to be dis
played horizontally. (Signals connected to
the — INPUT connector w ill be displayed
inverted.)
FOCUS
ASTIGMATISM
INTENSITY
POWER AND
SCALE ILLUM.
CAL. OUT con
nectors
HORIZONTAL
DISPLAY
POSITION
SENSITIVITY
SENSITIVITY
VARIABLE (red)
Focuses the trace or spot on the screen.
Controls roundness of the spot on the
screen. (This control is located inside on
instruments with serial numbers 000270
and above).
Controls the brightness of the trace or
spot on the screen.
Turns instrument power on and off, and
controls graticule illumination.
Provide amplitude-calibrated square waves
of 5 and 500 millivolts for use in calibrat
ing gain of amplifiers.
Selects conventional (horizontal sweep) or
X-Y mode of operation.
Also provides
magnification of sweep rate selected by
SWEEP TIME/CM control.
VERTICAL CHANNEL
Controls vertical positioning of the display
on the screen.
Selects the sensitivity of the Vertical Am
plifier from 14 calibrated steps.
Provides a continuous range of sensitivity
values between the fixed calibrated steps
selected by SENSITIVITY control.
(Nor
mally, sensitivity is calibrated only when
VARIABLE control is in CALIBRATED posi
tion.)
-(-INPUT and
Select desired coupling for incoming sig-
— INPUT
nal.
switches
PRELIMINARY INSTRUCTIONS
To set up the Type 503 Oscilloscope for operation, pro
ceed as follows:
1. Connect the oscilloscope to a suitable source of power
and turn the POWER switch clockwise from the POWER
OFF position. Set the INTENSITY control fully counterclock
wise. Let the instrument warm up for about 15 minutes.
2. Set the HORIZONTAL DISPLAY switch to the HORIZ.
AMPLIFIER position.
3. Set all of the INPUT switches to the G N D position.
4. Set the two SENSITIVITY switches to the .2 VOLTS/CM
position and the SENSITIVITY VARIABLE controls to the
CALIBRATED position (fully clockwise until they snap into
position).
5. Set both POSITION controls to midrange and advance
the INTENSITY control to produce a spot on the screen.
The INTENSITY control should be set to produce an easily
seen, but not bright, spot. Too bright a spot may damage
the phosphor on the face of the CRT. The intensity should
never be turned up to the point where a halo forms around
the spot.
6. Adjust the FOCUS control to produce the smallest,
roundest spot possible. Note the position of the spot on the
CRT.
©
2-2
Summary of Contents for 503
Page 5: ...Fig 1 1 Type 503 Oscilloscope Type 503 ...
Page 10: ...O perating Instructions Type 503 2 3 Fig 2 2 Type 503 Oscilloscope front panel ...
Page 23: ...NOTES ...
Page 33: ...NOTES ...
Page 49: ...Type 503 colibrotion__ 6 2 ...
Page 78: ...NOTES ...
Page 111: ...G Ab A B L O C K D I A G R A M ...
Page 112: ...T R K a G iE B IN P U T A M P L IF IE R T V p E 5 0 3 O SCILLO SCO PE D ...
Page 115: ... D Z 2 I j O H OA uit zul Z Q 3 O 0 7 7 ...
Page 116: ...c L O a l LLl A uJ u e x a uJ LLl t i ...
Page 118: ...0 S 2 J 7 ...
Page 119: ... J 0 o lii I in I I I 2 z 3i i pfSg D T IM IN G i S W IT C H ...
Page 121: ...n I u J a d i ...
Page 132: ...FIG 1 FRONT ...
Page 133: ...FIG 1 FRONT TYPE 503 OSCILLOSCOPE ...
Page 134: ...FIG 2 SWITCHES ...
Page 135: ...FIG 2 SWITCHES TYPE 503 OSCILLOSCOPE ...
Page 136: ...FIG 3 CRT SHIELD 2 5 A I ...
Page 137: ...FIG 3 CRT SHIELD f TYPE 503 OSCILLOSCOPE ...
Page 138: ...FIG 4 CHASSIS REAI ...
Page 139: ...TYPE 503 OSCILLOSCOPE ...
Page 140: ...FIG 5 CABINET 7 ...
Page 141: ...TYPE 503 OSCILLOSCOPE ...