See also
Other panel (Channel configuration menu)
on page 164
Deskew configuration menu (Other panel Channel configuration menu)
on page 164
Other panel (Channel configuration menu)
Use the Other panel to set the channel deskew, external attenuation, and alternate vertical scale units.
To open the analog Channel configuration menu Other panel:
1. Double-tap an analog Channel badge on the Settings bar to open the Channel configuration menu.
2. Tap the Other panel.
Other panel fields and controls
Not all items listed in the table are shown for all measurements; The panel only shows fields and controls relevant to the selected
measurement.
Field or control
Description
Deskew
Sets or displays the probe deskew value.
Use Deskew to make display and measurement adjustments for probes that have differing propagation
delays. This is especially important when using a current probe in conjunction with a voltage probe.
Set to 0
Sets the probe deskew value to zero (0) seconds.
Multi-Channel
Opens a Deskew configuration menu that allows you to deskew multiple channels (two at a time).
External Attenuation
Double-tap the numeric fields to set the external attenuation using the virtual keypad. Use this control
to set the input/output ratio of any external attenuation or gain between the signal and input channels.
Probe type
Selects the probe type.
Measure Current
Toggles between Yes and No. This control is only visible when Probe Type is set to Voltage.
Ratio
Sets the ratio using the virtual keypad. These fields show the conversion from volts to amps and amps
to volts. As you change one, the other changes as well. This control is only present when Measure
Current is set to Yes.
Set to Unity
Sets the external attenuation ratio to unity. Only shown when Alternate Units = On.
See also
Probe Compensation configuration menu (analog channels Probe Setup panel)
on page 163
Deskew configuration menu (Other panel Channel configuration menu)
on page 164
Deskew configuration menu (Other panel Channel configuration menu)
Use the Deskew configuration menu to make display and measurement adjustments for analog probes that have differing propagation
delays. This is especially important when using a current probe in conjunction with a voltage probe for power measurements.
To open the Deskew configuration menu:
1. Double-tap an analog Channel badge on the Settings bar to open the Channel configuration menu.
Menus and dialog boxes
3 Series Mixed Domain Oscilloscope Printable Help