CATV Measurements
5–30
2715 User Manual
9. Press [3] to print the results of the last measurement performed. A printer
must be connected to the 2715.
NOTE
. You can print measurement results only when the RS-232 interface is
installed in the 2715. The GPIB interface does not support the print function.
Be sure to select handshaking ([UTIL] [4] [0] [2] [5]) to match that expected by
the printer or long jobs may not print completely.
10. Press [4] to print all stored results of the measurement. A printer must be
connected to the 2715.
11. Press [9] to leave the Modulation Depth measurement and return to the
CATV MEASUREMENTS menu.
The modulation depth adjustment mode allows you to make real-time adjust-
ments of modulation depth. You select SET UP ADJUSTMENT MODE to
display the SET UP MOD DPTH ADJ MODE menu. Using this menu, you can
select whether or not you view a full field or a single line. You also use this
menu to set the target line, cycle delay, and target line duration. The SET UP
MOD DPTH ADJ MODE menu choices are listed below.
0
VIEW MODULATION MODE
FIELD (LINE)
1
TARGET LINE
87.5%
2
CYCLE DELAY
1000MSEC
3
TARGET LINE DURATION
30.0%
The VIEW MODULATION MODE provides two choices: FIELD or LINE. If
you select FIELD, you view a full field; if you select LINE, you will view line
#17 unless you have specified a VITS line for the channel in the channel table.
With TARGET LINE selected, you can set the target line to the depth of
modulation for which you want to adjust. The target line can be set to represent
values from 0% to 100%, with 87.5% the default setting.
With CYCLE DELAY selected, you can change the duration of the adjustment
cycle. The adjustment cycle consists of a display of the channel field (or line)
that alternates with a display of the target line and its associated readouts.
You can use the cycle delay with the target line duration (discussed in the
following text) to adjust the display times for comfortable viewing. The cycle
delay values can range from 0 ms to 30,000 ms.
With TARGET LINE DURATION selected, you can adjust how much of the
cycle delay is used to display the target line and its associated readouts. You can
use the target line duration with the cycle delay (discussed in the previous text)
to adjust the display times for comfortable viewing. The target line duration
values can range from 0% to 100%.
ADJUSTMENT MODE
Summary of Contents for 2715
Page 4: ......
Page 12: ...Table of Contents viii 2715 User Manual ...
Page 18: ...Preface xiv 2715 User Manual ...
Page 19: ...Introduction ...
Page 20: ......
Page 24: ...Introduction 1 4 2715 User Manual ...
Page 25: ...Getting Started ...
Page 26: ......
Page 46: ...Getting Started 2 20 2715 User Manual ...
Page 47: ...Operating Basics ...
Page 48: ......
Page 70: ...Operating Basics 3 22 2715 User Manual ...
Page 71: ...Dedicated Controls ...
Page 72: ......
Page 108: ...Dedicated Controls 4 36 2715 User Manual ...
Page 109: ...CATV Measurements ...
Page 110: ......
Page 181: ...Spectrum Analyzer Menus ...
Page 182: ......
Page 274: ...Spectrum Analyzer Menus 6 92 2715 User Manual ...
Page 275: ...Appendices ...
Page 276: ......
Page 302: ...Appendix B Broadcast AM FM and TV Signal Sources B 2 2715 User Manual ...
Page 318: ...Appendix C System Messages C 16 2715 User Manual ...
Page 334: ...Appendix E External Input and Output E 12 2715 User Manual ...
Page 339: ...Glossary and Index ...
Page 340: ......
Page 359: ......
Page 360: ......