Command Descriptions
3–32
Option 01 VXI Interface Module User Manual
Command
Response
trigger:sequence:
source hold
Disable all trigger sources.
initiate:continuous
Place the trigger subsystem in the arm state.
trigger:seq:imm
Override the hold placed on triggers. Close channel 1 in
section 1 of the VX4330.
trigger:seq:imm
Override the hold placed on triggers. Open channel 1 in section
1, then close channel 2 in section 1 of the VX4330.
trigger:seq:sour imm
Remove the hold placed on triggers. Set the trigger source to
immediate. This causes the sequencing of the scan list
indefinitely or until an ABORt command is received.
ABOR
Return the trigger subsystem to the idle state. Open the
currently closed relay in the scan list.
initiate:immediate
Place the trigger subsystem in the arm state, sequence through
the scan list twice, then return the trigger subsystem to the idle
state.