Operator Tutorial
2–10
1502C MTDR User Manual
O
F
F
O
F
F
O
F
F
O
N
ac
>=1 K
W
578.000 ft
Figure 2–12: Ohms-at-Cursor Beyond Reflected Pulse
9. Turn the
n
o
POSITION control to move the cursor to the far left side of the
display (–2.000 ft). Note that the ohms-at-cursor reading is now < 1
W
.
O
F
F
O
F
F
O
F
F
O
N
ac
< 1
W
–2.000 ft
Figure 2–13: Ohms-at-Cursor Beyond Reflected Pulse
If the cursor is placed too near a fault, the reflection will not have stabilized,
which will make the ohms-at-cursor reading misleading. This is especially true
very near the instrument where some aberrations are still significant. See the
Ohms-at-Cursor section of the Operating Instructions chapter for more on the
limitations of this feature.