Model 8717 Hypodermic Probe for the measurement of soft and semi-soft

materials. Miniature junction responds with one second time constant. All stain-

less steel construction is suitable for food industry, biological and chemical appli-

cations. Accuracy is ±2.2°C or 0.75%, whichever is greater. Time constant is

1sec (liquids). Max tip temperature is 500°C or 932°F.

Model 8731 Package of 5 miniature thermocouple connectors. Type K (male).

Model 8732 Standard to miniature thermocouple adapter. The female end of

the adapter is on the standard side. The male end of the adapter is on the minia-

ture side.

TYPE J PROBES (For Model 879)

Model 8722 Thermocouple Sensor is a three foot, 24 gauge thermocouple

wire that has a beaded thermocouple junction. Accuracy is ±2.2°C or 0.75%,

whichever is greater. Time constant is < 1sec. Max tip temperature is 480°C or

900°F.

Model 8723 Regular General Purpose/Immersion Probe for liquids and

general applications. Accuracy is ±2.2°C or 0.75°, whichever is greater. Time

constant is 1sec (stirred liquids). Max tip temperature is 760°C or 1400°F.

Model 8725 Regular Surface Probe for measurements of flat surfaces of

solids. Keeping the sensor tip flush to most surfaces provides fast responding

measurements. Accuracy is ±2.2°C or 0.75%, whichever is greater. Time con-

stant is 3sec (High thermal capacity/conductive solids). Max tip temperature is

400°C or 750°F.

Model 8726 Regular Air/Gas Probe for measurement of air/gasses. Exposed

junction within protection shroud responds to ambient temperature while shield-

ed from incident radiation. Accuracy is ±2.2°C or 0.75%, whichever is greater.

Time constrant is 3sec. (5m/s air flow). Max tip temperature is 300°C or 570°F.

TYPE T PROBES (For Models 875F & 875C)

Model 8752 Thermocouple Sensor is a three foot, 24 gauge thermocouple

wire sensor that has a beaded thermocouple junction. Accuracy is a ±2.2°C or

0.75%, whichever is greater. Time constant is <1sec. Max tip temperature is

400°C or 752°F.

Model 8753 Compact General Purpose/Immersion Probe for liquids and

general purpose applications. Has a one foot (5’ extended) coil cord, with a

molded plastic handle. Accuracy is a ±2.2°C or 0.75%, whichever is greater.

Time constant is 2sec (stirred liquids). Max tip temperature is 200°C or 392°F.

Model 8754 Compact Penetration Probe for measurements within soft or

semi-frozen materials. Has a one foot (5’ extended) coil cord, with a molded

plastic handle. Accuracy is ±2.2°C or 0.75%, whichever is greater. Time con-

stant is 1sec (stir-red liquids). Max tip temperature is 200°C or 392°F.

Model 8756 Compact Air/Gas Probe for measurement of air/gasses.

Exposed junction within protective shroud responds to ambient temperature

while shielded from incident radiation. Has a one foot (5’ extended) coil cord,

with a molded plastic handle. Accuracy is ±2.2°C or 0.75%, whichever is

greater. Time constant is 1.5sec (5 m/s air flow). Max tip temperature is 300°C

or 570°F.

8

SPECIFICATIONS

TEGAM

T.C.

TEMPERATURE

ACCURACY2

MODEL     TYPE1   RANGE      SPAN             RESOLUTION  (18° to 28°C; 1 year) 

874F        K        200°F   -40.0° to 199.9°F        0.1°        ±(0.5% rdg. + 1.0°F)

2000°F    -120° to 1999°F         1°           ±(1% rdg. + 2°F)

874C        K        200°C    -40.0° to 199.9°C   

0.1°        ±(0.5% rdg. + 0.5°C)

1100°C       -85° to 1100°C         1°          ±(1% rdg. + 1°C)

875F

T        200°F   -112.0° to 199.9°F       0.1°       ±(0.4% rdg. + 1.1°F)

752°F       200° to 752°F         

1°      

±(1% rdg. + 2°F)

875C        T        200°C    -80.0° to 199.9°C       0.1°      ±(0.4% rdg. + 0.6°C)

400°C      200° to 400°C           1°         ±(1% rdg. + 1°C)

878          K             °F     -120° to 1999°F           1°         ±(1% rdg. + 2°F)

°C      -85° to 1100°C           1°          ±(1% rdg. + 1°C)

879          J              °F     -112° to 1400°F       

1°          ±(1% rdg. + 2°F) 

°C       -80° to 760°C            1°          ±(1% rdg. + 1°C)

GENERAL SPECIFICATIONS

INPUT CURRENT:

< 1 microamp, typical.

REPEATABILITY:

0.3°F (0.2°C) typical for 1 week at constant ambient temperature.

TEMPERATURE COEFFICIENT:

18° to 28°C; included in accuracy specification.

From -10° to 18°, and 28° to 50°C; less than ±(0.025% of rdg +0.1°C)/°C.

ENVIRONMENTAL LIMITS FOR OPERATING:

-10° to 50°C, less than 80% rela-

tive humidity (R.H.) up to 35°C; reduce R.H. limit by 3%/°C from 35° to 50°C.

ENVIRONMENTAL LIMITS FOR STORAGE:

-35° to 60°C, less than 90% relative

humidity up to 35°C, reduce R.H. limit by 3%/°C from 35° to 60°C.

DISPLAY: 

3 1/2 digit LCD, 13mm (0.5”) height. Polarity indication. Decimal point when

req’d.

CONVERSION RATE:

1.3 readings per second.

OVERRANGE AND OPEN SENSOR INDICATION:

3 least significant digits

blanked.

MAXIMUM COMMON MODE VOLTAGE:

42V peak to earth.

THERMOCOUPLE INPUT:

Miniature TC Connector.

POWER: 

9V alkaline or carbon-zinc (NEDA 1604) battery.

BATTERY LIFE, CONTINUOUS: 

350 hours typical with alkaline; 200 hours typical

with carbon-zinc.

BATTERY INDICATOR: 

Display indicates “LO BAT” when less than 10% of life

remains.

DIMENSIONS: 

160mm long x 69mm wide x 31mm thick (6.3” x 2.7” x 1.2”).

Net weight 210gm (7.5 oz.)

CONSTRUCTION:

Heavy duty ABS plastic housing.

ACCESSORIES SUPPLIED: 

Battery and Instruction Manual.

Specifications subject to change without notice.

1Type K=NiCr-NiAl Thermocouple, Type T=Cu-CuNi Thermocouple, Type J=Fe-CuNi

Thermocouple.

2Excludes thermocouple sensor errors.

1

Summary of Contents for Budget 874F

Page 1: ...anufactured by Elcto Scientific Industries ESI Gertsch Keithley Instruments Lucas Weinschel and Pragmatic Instruments A complete list can be viewed on our Product Service Directory at www tegam com For more information about TEGAM and our products please visit our website at www tegam com or contact one of our customer service representatives at sales tegam com or 800 666 1010 Ten Tegam Way Geneva...

Page 2: ... to be defective To exercise this warranty write or call Tegam Inc in Geneva Ohio You will be given prompt assistance and return instruc tions Send the instrument transportation prepaid to the indi cated servive facility Repairs will be made and the instrument returned transportation prepaid Repaired products are war ranted for the balance of the original warranty period or at least 90 days LIMITA...

Page 3: ...n materials Has a one foot 5 extended coil cord with a molded plastic handle Accuracy is 2 2 C or 0 75 whichever is greater Time con stant is 1sec stir red liquids Max tip temperature is 200 C or 392 F Model 8756 Compact Air Gas Probe for measurement of air gasses Exposed junction within protective shroud responds to ambient temperature while shielded from incident radiation Has a one foot 5 exten...

Page 4: ...1A Thermocouple Kit includes the Model 8712 Thermocouple Sensor 6m 20 of AWG 24 Type K duplex TC wire a mininature TC plug and jack and a standard TC plug and jack Model 8712 Thermocouple Sensor is a three foot 24 gauge TC wire sensor that has a beaded thermocouple junction Accuracy is 2 2 C or 0 75 whichever is greater Time constant is 1sec Max tip temperature is 480 C or 900 F Model 8713 Regular...

Page 5: ...er similar surface and remove the screws from the bottom cover 2 Separate the bottom cover from the rest of the instrument by grasping the top of the case just above the display and care fully lifting it away from the display 3 Remove the old battery 4 Place the new battery in the battery compartment Be sure to observe the proper polarity Refer to Figure 2 5 Reinstall the bottom cover before resum...

Page 6: ...tion contains information needed to maintain your instrument The following information is included functional check performance veri fication and battery installation replacement WARNING The information presented in this section is intended for use by qualified personnel only Turn off the power and remove all probes from the instrument before replac ing the battery FUNCTIONALITY CHECK At room temp...

Reviews: